Verifying Performance

Agilent Technologies 8922M/S Specifications

Peak/Transmitter Carrier Power Measurement

RF In/Out only. After one timeslot, 577 μs, from an isolated Receiver Hop Trigger in the GSM bands.

Range

Input Frequency Setting Error

Input Level Setting Error

Accuracy

Supplemental Characteristics: Accuracy

Minimum Resolution

-5 to +41 dBm

±10 kHz

±3 dB

±0.6 dB ± noise effects (+4 to +41 dBm) (0.2 mW)

±0.6 dB ± noise effects (-5 to +4 dBm) (0.2 mW) 0.2 dB

Power Measurement Accuracy

Accuracy (dB)

0.5

0.4

0.3

0.2

0.1

0

-5

0

5

10

15

20

25

30

35

Power (dBm)

Pulse ON/OFF Ratio Measurement (Requires Option 006)

ON power is averaged over the useful part of the burst. OFF is averaged over a one bit interval centered at a user specified time. Non-hopped mode only.

Input Frequency Setting Error

Input Level Setting Error

Timing Accuracy

±10 kHz

±3 dB

±1.7 μs (± 1.1 μs typical)

3-19