7.When all readings are complete, examine the average DE* value. The instrument passes the repeatability test as long as the average DE* value is less than or equal to 0.04. Contact HunterLab Customer Support as described on page 25-6 if the instrument fails the repeatability test.

Note: Performance specifications are subject to change without notice.

Regulatory Notice

A copy of the Declaration of Conformity according to ISO/IEC Guide 22 and EN 45014 follows on the next page.