www.vxitech.com

INPUT:RANGE

The input range command selects the input range of one or more channels. The input range may be either set for ±10 volts or ± 100 volts. It is important to note that the input offset is normalized to ± 10 volt range. The actual input offset in the 100 volt range is ten times the set value.

INPut:RANGe <range>,<channel_list>

Where <range> is 10 V 100 V.

 

Where <channel_list> is the standard

 

channel list format supporting channels 1

 

through 16.

 

 

EXAMPLES

 

INPut:RANGe 100,(@1:16)

Sets the input range for Channels 1 through

 

16 to 100 V.

INPut:RANGe? 7

Returns the input range for Channel 7 as

100

100 V.

INP:RANG 10,(@4:6)

Sets the input range for Channels 4 through

 

6 to 10 V.

INP:RANG? 5

Returns the input range for Channel 5 as

10

10 V.

VM4016 Programming

31

Page 31
Image 31
VXI VM4016 user manual Inputrange