Agilent Technologies 1680 quick start Safety Notices, Warnings, Series Quick Start/Installation

Page 25

Safety Notices

This apparatus has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Measuring Apparatus, and has been supplied in a safe condition. This is a Safety Class I instrument (provided with terminal for protective earthing). Before applying power, verify that the correct safety precautions are taken (see the following warnings). In addition, note the external markings on the instrument that are described under "Safety Symbols."

Warnings

Before turning on the instrument, you must connect the protective earth terminal of the instrument to the protective conductor of the (mains) power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. You must not negate the protective action by using an extension cord (power cable) without a protective conductor (grounding). Grounding one conductor of a two-conductor outlet is not sufficient protection.

Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuse holders. To do so could cause a shock or fire hazard.

If you energize this instrument by an auto transformer (for voltage reduction or mains isolation), the common terminal must be connected to the earth terminal of the power source.

Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against any unintended operation.

Service instructions are for trained service personnel. To avoid dangerous electric shock, do not perform any service unless qualified to do so. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.

Do not install substitute parts or perform any unauthorized modification to the instrument.

Capacitors inside the instrument may retain a charge even if the instrument is disconnected from its source of supply.

1680-Series Quick Start/Installation

25

Image 25
Contents Agilent Technologies Changing Network SettingsMaking a Measurement Trigger HistoryFront Panel Product OverviewRear Panel 1680-Series Quick Start/Installation Accessories PouchGet Network Setup Information First-Time Power Up ConsiderationsDetermine Whether the Instrument will be Connected to the Network N O T EThe Workgroup or Domain name Domain Name Service DNSMethodology for assigning IP addresses at your site for example, DHCP Answering Windows Welcome Questions Answering Windows Welcome QuestionsStandalone No LAN WorkgroupChoose Yes, through a If expected Windows Welcome questions do not appearWindows unable to determine Ready to register with MicrosoftEstablishing an Administrator Account Additional First-Time Setup Stepsrecommended for standalone and networked instruments Protecting the Instrument’s Operating System2 Assign buses and signals and click OK Making a Measurement3 Enter a trigger value 4 Acquire and view data Snap to Edge Markers Tree Structure Labels Left-Click MenusShow individual signals in buses by clicking on the + to expand Marker Overview Offline AnalysisTool Tips Save and recall trigger settings Trigger HistoryProbing Changing Network SettingsTo give other applications/ports access through the firewall Changing Windows Firewall Settings3 In the Windows Firewall dialog, click the Exceptions tab To restore the logic analyzer firewall defaults 3 In the Windows Firewall dialog, click the Advanced tab Agilent Logic Analyzer Software Software Installation and UpgradesOptional Products Proper CoolingSelf-Test Emergency RecoveryElectrical Characteristics Specifications and CharacteristicsPower Environmental Characteristics Operating1 Click HelpHelp Topics 2 Click ReferenceSpecifications and CharacteristicsWarnings Safety NoticesSafety Symbols To clean the instrument1680-Series Quick Start/Installation Manual part number 01680-97012