Meade 114 EQ-D instruction manual Diagonal holder adjustments, Primary mirror adjustments

Page 13

– 13 –

To inspect the view of the mirror collimation, look down the focuser drawtube with the eyepiece removed. The edge of the focuser drawtube (1, Fig. 7), will frame the reflections of the primary mirror with the 3 mirror clips (2, Fig. 7), the diagonal mirror (3, Fig. 7) , the spider vanes (4, Fig. 7), and your eye (5, Fig. 7). Properly aligned, all of these reflections will appear concentric (i.e., centered) as illustrated in Fig. 7.

Any deviation from the concentric reflections will require adjustments to the diagonal assembly (Fig. 5), and/or the primary mirror cell (Fig. 6).

Remove adhesive backing

21

b. Diagonal holder adjustments

Fig. 5: Diagonal Assembly.

If the diagonal mirror (1, Fig. 8) is centered in the drawtube (2, Fig. 8), but the primary mirror is only partially visible in the reflection (3, Fig. 8), the 3 Phillips-head diagonal tilt screws (1, Fig. 5. Note: To adjust these screws you must first remove an adhesive backing) must be unthreaded slightly to the point of where you can tilt the diagonal holder (3, Fig. 5) from side-to-side by grasping the diagonal

holder with your hand and tilt until you see the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

primary mirror become as centered in the

 

 

 

 

 

 

 

 

 

 

 

 

 

reflection of the diagonal mirror as possible. Once

3

 

 

 

 

 

 

 

 

2

 

 

 

 

you are at the best position, thread in the 3

 

 

 

 

 

 

 

 

 

 

 

 

 

Phillips-head diagonal tilt screws to lock the

 

 

 

 

 

 

 

 

 

 

 

 

 

rotational position. Then, if necessary, make

 

 

 

 

 

 

 

 

 

 

 

 

 

adjustments to these 3 Phillips-head screws to

 

 

 

 

 

 

 

 

 

 

 

 

 

refine the tilt-angle of the diagonal mirror until the

 

 

 

 

 

 

 

 

 

 

 

 

 

entire primary mirror can be seen centered within

 

 

 

 

 

 

 

 

 

 

 

 

 

the diagonal mirror reflection. When the diagonal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

mirror is correctly aligned, it will look like Fig. 9.

 

 

 

 

 

 

 

 

 

 

 

 

 

(Note: the primary mirror is shown out of

 

 

 

 

 

 

 

 

 

 

 

 

 

alignment.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fig. 6: Primary Mirror Cell.

 

 

 

 

 

 

 

 

 

 

c. Primary mirror adjustments

If the diagonal mirror (1, Fig. 9) and the reflection of the primary mirror (2, Fig. 9) appear centered within the drawtube (3, Fig. 9), but the reflection of your eye and the reflection of the diagonal mirror (4, Fig. 9) appear off-center, you will need to adjust the primary mirror tilt Phillips-head screws of the primary mirror cell (3, Fig. 6). These primary tilt screws are located behind the primary mirror, at the lower end of the main tube. See Fig. 4. To adjust the primary mirror tilt screws, first unscrew several turns, the 3 hex-head primary mirror cell locking screws (2, Fig.6) that are next to each primary mirror tilt Phillips-head screw. Then by trial-and-error, turn the primary mirror tilt Phillips-head screws (3, Fig. 6) until you develop a feel for which way to turn each screw to center the reflection of your eye. Once centered, as in Fig. 7, turn the 3 hex-head primary mirror cell locking screws (2, Fig. 6) to relock the tilt-angle adjustment.

d. Star testing the collimation

With the collimation performed, you will want to test the accuracy of the alignment on a star. Use the MA 25mm eyepiece and point the telescope at a moderately bright (second or third magnitude) star, then center the star image in the telescope’s field-of-view. With the star centered follow the method below:

Bring the star image slowly out of focus until one or more rings are visible around the central disc. If the collimation was performed correctly, the central star disk and rings will be concentric circles, with a dark spot dead center within the out-of-focus star disk (this is the shadow of the secondary mirror), as shown in Fig. 10C. (An improperly aligned telescope will reveal elongated circles (Fig. 10A), with an off-center dark shadow.)

If the out-of-focus star disk appears elongated (Fig. 10A), you will need to adjust the primary mirror Phillips-head tilt screws of the primary mirror cell (3, Fig. 6).

To adjust the primary mirror tilt screws (3, Fig. 6), first unscrew several turns the 3 hex-head primary mirror cell locking screws (2, Fig. 6), to allow free turning movement of the tilt knobs.

Using the flexible cable controls (3 and 4, Fig. 1), move the telescope until the star image is at the edge of the field-of-view in the eyepiece, as in Fig. 10B.

Image 13
Contents Meade Instruments Corporation Meade Limited Warranty Table of Contents Key to Fig Reverse Introduction Unpacking and AssemblyThis Manual Standard EquipmentAlignment of the Viewfinder Balancing the TelescopeCelestial Sphere Understanding Celestial Movements and CoordinatesLining UP with the Celestial Pole Using the Telescope Calculating Power Using Setting CirclesCleaning MaintenanceMount and Tripod Adjustments Collimation Alignment of the OpticsPrimary mirror adjustments Diagonal holder adjustmentsStar testing the collimation ABC Optional Accessories SpecificationsMeade Instruments Corporation