QED6 Users | Manual Supplement |
Change #1, 56600
On page
Output Power Measurement
Load Resistance | 50 Ω ±1 % |
| (< 10 µH), 160 W |
Range |
|
400 J | |
100 J | |
Resolution | 0.1 J |
Max. Vage |
|
400 J | 5500 V |
100 J | 1750 V |
Max. Current |
|
400 J | 275 A |
100 J | 35 A |
Measurement | 400 J: 66 ±5 V |
Trip Levels | 100 J: 20 ±5 V |
Pulse Width | |
Accuracy |
|
400 J Range | ±2 % of reading |
±2 Js | |
100 J Range | ±2 % of reading |
| ±0.1 J |
Change #2
On page
•Select the 400 joule high range (400 J) for defibrillator outputs over 100 joules or for an unknown defibrillator output power.
On page
•Select the 400 joule high range for defibrillator outputs over 100 joules or for an unknown defibrillator output power.
11/10 | 1 |