Fluke 2204520 manual Change #1, Change #2, Output Power Measurement

Page 2

QED6 Users

Manual Supplement

Change #1, 56600

On page 1-10, replace the Output Power Measurement with the following:

Output Power Measurement

Load Resistance

50 Ω ±1 % non-inductance

 

(< 10 µH), 160 W

Range

 

400 J

0-400.0 J

100 J

0-100.0 J

Resolution

0.1 J

Max. Vage

 

400 J

5500 V

100 J

1750 V

Max. Current

 

400 J

275 A

100 J

35 A

Measurement

400 J: 66 ±5 V

Trip Levels

100 J: 20 ±5 V

Pulse Width

1-50 ms

Accuracy

 

400 J Range

±2 % of reading

100-400 J

±2 Js

100 J Range

±2 % of reading

 

±0.1 J

Change #2

On page 2-10, under the second step 4, replace the first bullet with the following:

Select the 400 joule high range (400 J) for defibrillator outputs over 100 joules or for an unknown defibrillator output power.

On page 2-17, under step 5, replace the first bullet with the following:

Select the 400 joule high range for defibrillator outputs over 100 joules or for an unknown defibrillator output power.

11/10