Victoreen 
Operators Manual
1.2 Specifications
Detector  | 
  | |||
Energy Dependence  | 10% to 40% from 50 KeV to 2 MeV, as referenced to 137Cs  | |||
Overload Protection | Alarms in fields higher than 100 R/h when detector is completely blocked for  | |||
  | pulse operation.  | 
  | ||
Operation Indicator  | Provide continuous visual indication of background radiation.  | |||
Range/lndicators  | 1  | mR/h: Green  | 
  | |
  | 2  | mR/h: Green  | 
  | |
  | 4  | mR/h: Yellow  | 
  | |
  | 8  | mR/h: Yellow  | 
  | |
  | 16 mR/h: Red  | 
  | ||
  | 32 mR/h: Red  | 
  | ||
Response Time  | Level  | Response Time  | 
  | |
  | 1 mR/h  | 3 seconds  | 
  | |
  | 2 mR/h  | 1.5 seconds  | 
  | |
  | 4 mR/h  | 800 milliseconds  | 
  | |
  | 8 mR/h  | 400 milliseconds  | 
  | |
  | 16 mR/h  | 200 milliseconds  | 
  | |
  | 32 mR/h  | 100 milliseconds  | 
  | |
Alarms | Alarm may be preset by user at any indicated discrete level by means of the  | |||
  | front panel selector switch; front panel switch selects all alarms or visual  | |||
  | only.  | 
  | 
  | |
  | Visual Alarm: bright incandescent red lamps flashing on/off approximately  | |||
  | 1/sec.  | 
  | 
  | |
  | Aural Alarm: 400 Hz tone, alternating 1/sec; sound level = 76 db at one  | |||
  | foot.  | 
  | 
  | |
Operating Conditions  | 10° C to 40° C (50° F to 104° F)  | 
  | ||
  | Maximum of 90% relative humidity   | |||
External Connectors | Jack located on case bottom allows attachment  | of external aural/visual  | ||
  | alarm (Primalarm) at remote location.  | 
  | ||
Power Supply | Part   | 
  | ||
  | 
  | Input 117 VAC, 60 Hz, 12 W  | 
  | |
  | 
  | Output 12 VDC, 500 mA  | 
  | |
Part 
Input 230 VAC, 50 Hz,
Output 12 VDC, 500 mA
Part