Victoreen
Operators Manual
1.2 Specifications
Detector |
| |||
Energy Dependence | 10% to 40% from 50 KeV to 2 MeV, as referenced to 137Cs | |||
Overload Protection | Alarms in fields higher than 100 R/h when detector is completely blocked for | |||
| pulse operation. |
| ||
Operation Indicator | Provide continuous visual indication of background radiation. | |||
Range/lndicators | 1 | mR/h: Green |
| |
| 2 | mR/h: Green |
| |
| 4 | mR/h: Yellow |
| |
| 8 | mR/h: Yellow |
| |
| 16 mR/h: Red |
| ||
| 32 mR/h: Red |
| ||
Response Time | Level | Response Time |
| |
| 1 mR/h | 3 seconds |
| |
| 2 mR/h | 1.5 seconds |
| |
| 4 mR/h | 800 milliseconds |
| |
| 8 mR/h | 400 milliseconds |
| |
| 16 mR/h | 200 milliseconds |
| |
| 32 mR/h | 100 milliseconds |
| |
Alarms | Alarm may be preset by user at any indicated discrete level by means of the | |||
| front panel selector switch; front panel switch selects all alarms or visual | |||
| only. |
|
| |
| Visual Alarm: bright incandescent red lamps flashing on/off approximately | |||
| 1/sec. |
|
| |
| Aural Alarm: 400 Hz tone, alternating 1/sec; sound level = 76 db at one | |||
| foot. |
|
| |
Operating Conditions | 10° C to 40° C (50° F to 104° F) |
| ||
| Maximum of 90% relative humidity | |||
External Connectors | Jack located on case bottom allows attachment | of external aural/visual | ||
| alarm (Primalarm) at remote location. |
| ||
Power Supply | Part |
| ||
|
| Input 117 VAC, 60 Hz, 12 W |
| |
|
| Output 12 VDC, 500 mA |
|
Part
Input 230 VAC, 50 Hz,
Output 12 VDC, 500 mA
Part