4 Product Specifications
Therefore,
ETHERMOCOUPLE = 0.75% × 600 °C = 4.5 °C
Using the above individual component errors, the total measurement system accuracy is calculated as below.
System accuracy without thermocouple sensor error:
ETOTAL
=SQRT(EDAQ2 + ESIG_COND2 + EZEROING2 + ECJC2)
=0.95 °C
System accuracy with thermocouple sensor error:
ETOTAL
=SQRT(EDAQ2 + ESIG_COND2 + EZEROING2 + ECJC2 + ETHERMOCOUPLE2)
=4.6 °C
56 | Agilent U2802A |