Option Installation
Update Procedure
Writing System Correction Data of Backup Memory to Removable
HDD
The following is a description of the procedures to write the system correction data stored in the backup memory to the mounted removable HDD.
Step 1. Turn on the E5070B/E5071B.
Step 2. | Press | . |
Step 3. Press Load Project... on the softkey menu. | ||
Step 4. A dialog box appears to allow you to select a program to be loaded. Select | ||
| RestoreSysCorFile.vba from the D:\Agilent\Service folder, and then press Open button. | |
Step 5. | Press | . |
Step 6. | RestoreSysCorFile dialog box appears (Figure | |
Figure | RestoreSysCorFile Dialog Box |
CAUTION | Once the program has ended, restart the E5070B/E5071B to enable the system correction |
| data. |
8 | Chapter 2 |