TROUBLESHOOTING GUIDE
Trouble Shooting Guide
In the event of an error code, unplug the Attest 290
The display panel will indicate malfunctions detected by a specific error code. Refer to the following:
Code | Problem | Solution |
C1 Unit
Allow a 30 minute
C2 RRBI has been removed before incubation Replace RRBI in proper well within 10 seconds. complete. All three lights for affected RRBI
will light and alarm will sound if active.
E1 | Temperature Control Error |
| This error will occur if system is no |
longer able to control the temperature of the incubator block (60ºC
Call for repair or replacement.
E3 | Memory Error. | Call for repair or replacement. |
E4 | Call for repair or replacement. | |
E7 | Communications Error | Call for repair or replacement. |
| This error will occur if internal |
|
| communications have failed. |
|
E8 | Bulb Error |
| This error occurs if the system detects |
| that the bulb is too dim to properly read |
| RRBI results. |
Call for repair or replacement.
E9 | Transport Error |
| This error will occur if the system detects |
| a motor failure or positioning problem. |
Call for repair or replacement.
3M © 2003 | Page 8 of 9 | Attest 290 |