Agilent Technologies Scan Feature, Single Scan List Size, Scan Rate 3499A/B/C, Modules Ch/s

Page 8

Scan Feature

The scan feature is a sequence of alternating switch closures and measurements. This feature is provided by both the 34980A and a 3499A/B/C combined with a DMM or other measurement instrument. The 34980A increases the measurement scan abilities and offers more than twice the number of scanned channels within one scan list compared to the 3499A/B/C as seen here.

Single Scan List Size:

34980A

Up to 560

 

2-wire channels

 

or

 

Up to 640

 

1-wire channels

3499A/B/C

Up to 200 channels

 

and/or bits

Although the 34980A scan feature can be used with either the internal DMM or with an external instrument, using the integrated DMM provides a faster and more efficient scan rate. The integrated DMM eliminates the need for external instrument triggers and wiring. The 3499A/B/C scan list requires the use of instrument triggers to work with an external DMM or other measurement instrument. Each switch module is rated with individual switch scan rates. A comparison of the scan rates for both systems Multiplexer and Matrix switch modules is shown in Table 5 below.

Another advantage the 34980A system has over the 3499A/B/C system is its ability to store measurement readings. When the 34980A internal DMM is used, up to 500,000 time stamped readings can be stored in volatile memory.

The 3499A/B/C cannot store readings but does offer more scan configurability. For example, once a 3499A/B/C scan list is set up, the user can select an arm source, a trigger source, the number of sweeps, and the delay time for each individual channel within the scan list.

Table 5. 34980A and 3499A/B/C Multiplexer & Matrix module scan rates

 

 

 

 

Module Type

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

34980A

 

Scan Rate

3499A/B/C

Scan Rate

 

 

Modules

Ch/s

Modules

Ch/s

 

Multiplexer

34921A 40 Ch armature

100

N2260A 40 Ch armature

80

 

 

34922A

70 Ch armature

100

N2266A 40 Ch reed

350

 

 

34923A

40 Ch reed

500

N2270A 10 Ch

100

 

 

34924A

70 Ch reed

500

44470A

10 Ch armature

43

 

 

34925A

40 Ch FET

1000

44470D

20 Ch armature

43

 

 

 

 

 

 

 

Matrix

34931A Dual 4x8 armature

100

N2262A 4x8 armature

80

 

 

34932A Dual 4x16 armature

100

44473A

4x4 armature

43

 

 

34933A Dual 4x8 reed

500

 

 

 

8

Image 8
Contents Application Note MeasurementsChannels 2-wire MainframesMainframe Maximum NumberModules Benefits of the new 34980A modules vs 3499A/B/C modulesModule Type 3499A/B/CUser Interface Instrument Control ProgrammingWeb Page Interface Scpi Commands 34980AScan Feature Single Scan List SizeScan Rate 3499A/B/C Modules Ch/sMonitor Feature Stored State FeatureSwitch and Measurement Connections 3499A/B/C differentiating features SummaryAppendix a Command Reference/Comparison 34980A Commands 3499A/B/C CommandsARM 01OFFON34980A Commands cont’d Data LAST?SPEEK? SpokeREAD? Open OPEN? Openabus OpenallScan Scanadd SCANSIZE? SCANLIST? ScanlistDATABIT? DATABYTEWORDLWORDBLOCK?Type FrtdtypeRtdtype DATA?DCNPLC? DczeroautoDCZEROAUTO? ModeALARM? Landhcp LandnsLANMAC? CponTIMESCAN? Data DatadacFREE? UnitOur Promise