Contents
Agilent 34970A Data Acquistion/Switch Unit
User’s Guide
Page
Convenient Data Logging Features
Flexible Data Acquisition/Switching Features
Front Panel at a Glance
Front-Panel Menus at a Glance
Display Annunciators
Adrs
Rear Panel at a Glance
Use the Menu to
BenchLink Data Logger Software at a Glance
For detailed information and a module diagram, see
Plug-In Modules at a Glance
34901A 20-Channel Armature Multiplexer
34902A 16-Channel Reed Multiplexer
34904A 4x8 Two-Wire Matrix Switch
34903A 20-Channel Actuator/General-Purpose Switch
34905/6A Dual 4-Channel RF Multiplexers
34907A Multifunction Module
For detailed information and module block diagrams, see
34908A 40-Channel Single-Ended Multiplexer
This Book
Contents
Contents
Remote Interface Reference
Application Programs
Quick Start
Quick Start
Connect the power cord and turn on the instrument
To Prepare the Instrument for Use
To Prepare the Instrument for Use
Check the list of supplied items
Installation Procedure
Installing BenchLink Data Logger Software
Installing BenchLink Data Logger Software
Creating Installation Floppy Disks
On-Line Help System
To Connect Wiring to a Module
To Connect Wiring to a Module
Wire Ohms / RTD / Thermistor
Wire Ohms / RTD
Thermocouple DC Voltage / AC Voltage / Frequency
DC Current / AC Current
To Set the Time and Date
To Set the Time and Date
Set the time of day
Set the date
To Configure a Channel for Scanning
To Configure a Channel for Scanning
Select the channel to be added to the scan list
Select the measurement parameters for the selected channel
Run the scan and store the readings in non-volatile memory
View the data from the scan
To Copy a Channel Configuration
To Copy a Channel Configuration
To Close a Channel
To Close a Channel
Select the channel
Close the selected channel Open the selected channel
Verify that there is ac power to the instrument
Verify the power-line voltage setting
Verify that the power-line fuse is good
If the Instrument Does Not Turn On
Replace the fuse-holder assembly in the rear panel
To Adjust the Carrying Handle
To Adjust the Carrying Handle
Bench-top viewing positions
To Rack Mount the Instrument
To Rack Mount the Instrument
Quick Start
Page
Front-Panel Overview
Front-Panel Overview
Front-Panel Menu Reference
Front-Panel Menu Reference
Configure system-related instrument parameters
Enable monitoring on the selected channel
To disable monitoring, press again
To Monitor a Single Channel
To Monitor a Single Channel
Select the interval scan mode
To Set a Scan Interval
To Set a Scan Interval
Select the scan count
To Apply Mx+B Scaling to Measurements
To Apply Mx+B Scaling to Measurements
To Configure Alarm Limits
To Configure Alarm Limits
Select which of the four alarms you want to use
Select the alarm mode on the selected channel
Set the limit value
To Read a Digital Input Port
To Read a Digital Input Port
Select the Digital Input port
Read the specified port
To Write to a Digital Output Port
To Write to a Digital Output Port
Configure the totalize mode
To Read the Totalizer Count
To Read the Totalizer Count
Select the totalizer channel
To Output a DC Voltage
To Output a DC Voltage
Select a DAC Output channel
Output the voltage from the selected DAC
To Configure the Remote Interface
To Configure the Remote Interface
Select the Gpib Hpib interface
Select the Gpib address
Select the RS-232 interface Select the baud rate
Select the parity and number of data bits
Select the flow control method
Save the changes and exit the menu
To Store the Instrument State
To Store the Instrument State
Select the storage location
Store the instrument state
System Overview
Data Acquisition System Overview
Control Output, starting on
Data Acquisition System Overview
Computer and Interface Cable
Advantages Disadvantages
Measurement Software
Agilent BenchLink Data Logger
34970A Data Acquisition / Switch Unit
Plug-In Modules
Signal Routing
Model Number Module Name Common Uses
System Cabling and Connections starting on
System Cabling
Cable Type Common Uses Comments
Alarm Limits
Transducers and Sensors
Switching Topologies
Signal Routing and Switching
Signal Routing and Switching
Multiplexers are available in several types
Channel Open
Internal DMM
Measurement Input
Measurement Input
Measurement Input
Scanning
Scan Count
Scanning With External Instruments
Multifunction Module
+IN
Control Output
Control Output
DAC
Actuator / General-Purpose Switch
Page
Features and Functions
Mx+B Scaling, starting on Alarm Limits, starting on
Scanning, starting on
Scpi Language Conventions
Rules for Using a Channel List
Scpi Language Conventions
Rules for Scanning
Scanning
Scanning
Scanning
Or disable it see Internal DMM Disable on
Power Failure
Adding Channels to a Scan List
To Build a Scan List From the Front Panel
To Build a Scan List From the Remote Interface
Scan Interval
Select the interval timer configuration
See Scan Count on page 86 for more information
Set the scan interval to 5 seconds
Sweep the scan list 2 times
Front-Panel Operation
Select the bus once configuration
Group Execute Trigger
Ext Trig Connector
Select the external trigger configuration
On page 86 for more information
Select the alarm configuration
Enable the upper limit
Report alarms on Alarm
Enable monitoring
Scan Count
Remote Interface Operation
Reading Format
Channel Delay
Automatic Channel Delays
Integration Time Channel Delay
Range Channel Delay
AC Filter Channel Delay
Viewing Readings Stored in Memory
Readings
Minimum reading on channel
Time minimum was logged
Maximum reading on channel
Time maximum was logged
Single-Channel Monitoring
Single-Channel Monitoring
ROUTMONDATA?
Scanning With External Instruments
Scanning With External Instruments
Scanning With External Instruments
Select the scan interval
Select the channel advance source
General Measurement Configuration
General Measurement Configuration
Measurement Range
MV Range
Measurement Resolution
Integration Time, on page 103 for more information
102
Custom A/D Integration Time
PLC
104
Autozero
Temperature Measurement Configuration
Temperature Measurement Configuration
Measurement Units
Thermocouple Measurements
108
Remote Interface Operation You can use the MEASure? or
RTD Measurements
To connect an RTD to the module’s screw terminals, see
111
Thermistor Measurements
To connect a thermistor to the module’s screw terminals, see
Voltage Measurement Configuration
Voltage Measurement Configuration
DC Input Resistance
Applies to dc voltage measurements only
Applies to ac voltage and ac current measurements only
AC Low Frequency Filter
Sensvoltacband 3,@203 Select the slow filter 3 Hz
Resistance Measurement Configuration
Resistance Measurement Configuration
Offset Compensation
To connect resistances to the module’s screw terminals, see
Current Measurement Configuration
Current Measurement Configuration
Applies to ac current and ac voltage measurements only
Senscurracband 3,@221 Select the slow filter 3 Hz
Frequency Measurement Configuration
Low Frequency Timeout
Frequency Measurement Configuration
Sensfreqranglow 3,@203 Select the slow timeout 3 Hz
Mx+B Scaling
Mx+B Scaling
= − GF x R
121
Alarm Limits
Alarm Limits
Alarm Limits
Alarm Event No Alarm Upper Limit Lower Limit
Then, choose from the following alarm conditions
Viewing Stored Alarm Data
127
Using the Alarm Output Lines
Alarms Connector
Clear alarm output line
Clear all four alarm outputs
Using Alarms With the Multifunction Module
Following commands also see the example on the following
Calccompdata command decimal
Calccompmask command decimal
Result no alarm generated
Data read from port decimal
Digital Input Operations
Digital Input Operations
Read both ports together
Read port
Add port 02 read to scan list
Totalizer Operations
Totalizer Operations
136
137
Digital Output Operations
Digital Output Operations
Write to port
Write to both ports
DAC Output Operations
DAC Output Operations
Three slots six DAC channels
State Storage
System-Related Operations
System-Related Operations
141
Error Conditions
Self-Test
Returns 0 if the self-test is successful or 1 if it fails
Display Control
Internal DMM Disable
Real-Time System Clock
Set time to 345 PM
Set date to June 1
Firmware Revision Query
HEWLETT-PACKARD,34970A,0,X.X-Y.Y-Z.Z
Relay Cycle Count
148
Scpi Language Version Query
You cannot query the Scpi version from the front panel
Remote Interface Configuration
Remote Interface Configuration
Gpib Address
You can set the Gpib address from the front panel only
Remote Interface Selection
Baud Rate Selection RS-232
Parity Selection RS-232
You can set the baud rate from the front panel only
You can set the parity from the front panel only
Flow Control Selection RS-232
154
Calibration Security
Calibration Overview
Calibration Overview
156
Unsecure with old code
Enter new code
Calibration Message
Calibration Count
Factory Reset State
Factory Reset State
Instrument Preset State
Instrument Preset State
Multiplexer Module Default Settings
Multiplexer Module Default Settings
Module Overview
Module Overview
34901A 20-Channel Multiplexer
34901A 20-Channel Multiplexer
Slot Number
Name Function Comments
34902A 16-Channel Multiplexer
34902A 16-Channel Multiplexer
167
34903A 20-Channel Actuator
34903A 20-Channel Actuator
169
34904A 4x8 Matrix Switch
34904A 4x8 Matrix Switch
Name Comments
34905A/6A Dual 4-Channel RF Multiplexers
34905A/6A Dual 4-Channel RF Multiplexers
173
34907A Multifunction Module
Digital Input/Output
Totalize Input
Analog Output DAC
Slot Number 100 200
Digital Input / Output
Totalizer
DAC Output
34908A 40-Channel Single-Ended Multiplexer
34908A 40-Channel Single-Ended Multiplexer
177
178
Remote Interface Reference
Calibration Commands, starting on
Totalizer Commands, starting on
Scanning Overview, starting on
Scpi Command Summary
Scpi Command Summary
Rules for Using scanlist and chlist Parameters
Scpi Command Summary Scan Measurement Commands
Monitor Commands
See page 226 for more information
See page 237 for more information
Scpi Command Summary Scan Configuration Commands
FORMat
Scpi Command Summary Scan Statistics Commands
Scan Memory Commands
See page 233 for more information
See page 235 for more information
Scpi Command Summary Scanning With an External Instrument
See page 239 for more information
Scpi Command Summary Temperature Configuration Commands
See page 219 for more information
Scpi Command Summary Voltage Configuration Commands
See page 223 for more information
Scpi Command Summary Resistance Configuration Commands
See page 224 for more information
Scpi Command Summary Current Configuration Commands
See page 214 for more information
Scpi Command Summary Mx+B Scaling Commands
See page 244 for more information
Scpi Command Summary Alarm Limit Commands
See page 247 for more information
Scpi Command Summary Digital Input Commands
Totalizer Commands
See page 255 for more information
See page 256 for more information
Scpi Command Summary Digital Output Commands
DAC Output Commands
Switch Control Commands
See page 258 for more information
Scpi Command Summary Scan Triggering Commands
State Storage Commands
See page 228 for more information
See page 261 for more information
Scpi Command Summary System-Related Commands
See page 264 for more information
Scpi Command Summary Interface Configuration Commands
Status System Commands
See page 269 for more information
See page 286 for more information
Scpi Command Summary Calibration Commands
Service-Related Commands
See page 292 for more information
See page 294 for more information
Scpi Command Summary Ieee 488.2 Common Commands
Simplified Programming Overview
Simplified Programming Overview
Using the MEASure? Command
Using the CONFigure Command
Using the range and resolution Parameters
Using the READ? Command
Using the INITiate and FETCh? Commands
205
Trigsour EXT Init FETC?
MEASure? and CONFigure Commands
MEASure? and CONFigure Commands
MEASure? Command Syntax
MEASureTEMPerature?5 TCouple,BEJKNRSTDEF
MEASureTEMPerature? THERmistor,2252500010000DEF
MEASureTEMPerature? RTDFRTD,8591DEF
MEASureVOLTageDC? MEASureVOLTageAC? RangeAUTOMINMAXDEF
MEASureCURRentDC? MEASureCURRentAC? RangeAUTOMINMAXDEF
MEASureRESistance? MEASureFRESistance? RangeAUTOMINMAXDEF
On the 34901A multiplexer module
MEASureTOTalize? READRRESet ,@scanlist
MEASureDIGitalBYTE? @scanlist
MEASure? and CONFigure Commands CONFigure Command Syntax
CONFigureTEMPerature TCouple,BEJKNRSTDEF
CONFigureTEMPerature RTDFRTD,8591DEF
CONFigureTEMPerature THERmistor,2252500010000DEF
CONFigureRESistance CONFigureFRESistance RangeAUTOMINMAXDEF
CONFigureVOLTageDC CONFigureVOLTageAC RangeAUTOMINMAXDEF
CONFigureCURRentDC CONFigureCURRentAC RangeAUTOMINMAXDEF
CONFigureFREQuency CONFigurePERiod RangeAUTOMINMAXDEF
CONFigureDIGitalBYTE @scanlist
CONFigureTOTalize READRRESet ,@scanlist
CONFigure? @chlist
Setting the Function, Range, and Resolution
Setting the Function, Range, and Resolution
See also General Measurement Configuration in starting on
SENSeFUNCtion function,@chlist
Setting the Function, Range, and Resolution
Setting the Function, Range, and Resolution
Setting the Function, Range, and Resolution
Setting the Function, Range, and Resolution
Temperature Configuration Commands
General Temperature Commands
SENSeTEMPeratureTRANsducer
SENSeTEMPerature Nplc 0.020.2121020100200MINMAX,@chlist
Temperature Configuration Commands Thermocouple Commands
SENSeTEMPeratureRJUNction? @ch list
Temperature Configuration Commands RTD Commands
Thermistor Commands
Voltage Configuration Commands
See also Voltage Measurement Configuration in starting on
INPut
SENSe Zeroauto OFFONCEON,@chlist ZEROAUTO? @chlist
Current Configuration Commands
Resistance Configuration Commands
See also Current Measurement Configuration in starting on
Frequency Configuration Commands
Frequency Configuration Commands
See also Frequency Measurement Configuration in starting on
Scanning Overview
Scanning Overview
See also Scanning in starting on
Scan Interval
Scanning Overview Scanning Commands
TRIGger SOURce BUSIMMediateEXTernalALARm1234TIMer SOURce?
ROUTe Scan @scanlist SCAN?
ROUTeSCANSIZE?
ROUTe CHANnelDELay seconds,@ch list CHANnelDELay? @chlist
TRIGger TIMer secondsMINMAX TIMer?
TRIGger COUNt countMINMAXINFinity COUNt?
ABORt
ROUTe
INITiate
FORMat READingALARm Offon READingALARm?
Scanning Overview Reading Format Commands
FORMat READingCHANnel Offon READingCHANnel?
FORMat READingUNIT Offon READingUNIT?
FORMat READingTIME Offon READingTIME?
FORMat READingTIMETYPE ABSoluteRELative READingTIMETYPE?
Scanning Overview Scan Statistics Commands
CALCulateAVERageMINimum? @chlist
CALCulateAVERageMINimumTIME? @chlist
CALCulateAVERageMAXimum? @chlist
CALCulateAVERageAVERage? @chlist
CALCulateAVERagePTPeak? @chlist
CALCulateAVERageCOUNt? @chlist
CALCulateAVERageCLEar @chlist
DATAPOINts?
Scanning Overview Scan Memory Commands
DATAREMove? numrdgs
SYSTemTIMESCAN?
See Reading Format Commands on
FETCh?
Single-Channel Monitoring Overview
Single-Channel Monitoring Overview
ROUTe MONitorSTATe Offon MONitorSTATe?
ROUTe MONitor @channel MONitor?
ROUTeMONitorDATA?
Scanning With an External Instrument
TRIGger SOURce BUSIMMediateEXTernalTIMer SOURce?
TRIGger COUNt valueMINMAXINFinity COUNt?
Scanning With an External Instrument
INSTrumentDMMINSTalled?
INSTrument
Mx+B Scaling Overview
Mx+B Scaling Overview
See also Mx+B Scaling in starting on
Mx+B Scaling Overview Mx+B Scaling Commands
CALCulate
CALCulateSCALeOFFSetNULL @chlist
Alarm System Overview
Alarm System Overview
See also Alarm Limits in starting on
Alarm System Overview
Calclimitupper MAX,@101LOWER 9,@101LOWERSTATE on
OUTPut ALARm1234SOURce @chlist ALARm1234SOURce?
Alarm System Overview Alarm Limit Commands
CALCulate LIMitUPPer value,@ch list LIMitUPPer? @chlist
SYSTemALARm?
CALCulate LIMitLOWer value,@ch list LIMitLOWer? @chlist
OUTPut ALARmMODE LATChTRACk ALARmMODE?
Alarm System Overview Alarm Output Commands
OUTPut ALARmSLOPe NEGativePOSitive ALARmSLOPe?
Digital I/O Alarm Commands
COMPareDATA? @chlist
COMPareMASK mask,@chlist COMPareMASK? @chlist
See also Digital Input Operations in starting on
Digital Input Commands
SENSeDIGitalDATABYTEWORD? @chlist
SENSe TOTalizeTYPE READRRESet,@chlist TOTalizeTYPE? @chlist
Totalizer Commands
See also Totalizer Operations in starting on
SENSeTOTalizeCLEarIMMediate @chlist
Commands see Reading Format Commands on
SENSeTOTalizeDATA? @chlist
Digital Output Commands
DAC Output Commands
SOURceDIGitalSTATe? @chlist
SOURce
Switch Control Commands
SYSTemCPON 100200300ALL
ROUTeDONE?
This is equivalent to pressing from the front panel
SAV
State Storage Commands
RCL
MEMorySTATe Name 12345 ,name NAME?
MEMorySTATeDELete
MEMorySTATeVALid?
RECallAUTO?
MEMoryNSTates?
System-Related Commands
See also System-Related Operations in starting on
SYSTemDATE yyyy,mm,dd
SYSTemDATE?
IDN?
SYSTemCTYPe?
DISPlay
DISPlay Offon DISPlay?
DISPlayTEXTCLEar
SYSTemPRESet
INSTrument DMM Offon
See for a complete listing of the 34970A error messages
SYSTemERRor?
SYSTemVERSion?
Interface Configuration Commands
RS-232 Configuration Overview
RS-232 Interface Configuration
RS-232 Interface Configuration
RS-232 Flow Control Modes
RS-232 Data Frame Format
Connection to a Computer or Terminal
RS-232 Troubleshooting
Modem Communications
Modem Communications
For more information, see Flow Control Selection on
S0=1
What is a Condition Register?
What is an Event Register?
What is an Enable Register?
Scpi Status System
Agilent 34970A Status System
Status Byte Register
Bit Definitions Status Byte Register
Decimal Bit Number
Definition
Using Service Request SRQ and Serial Poll
Using *STB? to Read the Status Byte
To Interrupt Your Bus Controller Using SRQ
Using the Message Available Bit MAV
To Determine When a Command Sequence is Completed
Questionable Data Register
Bit Definitions Questionable Data Register
Scpi Status System
Standard Event Register
Bit Definitions Standard Event Register
Scpi Status System
Alarm Register
Bit Definitions Alarm Register
Standard Operation Register
Bit Definitions Standard Operation Register
Status Byte Register Commands
Status System Commands
See the table on page 277 for the register bit definitions
Questionable Data Register Commands
See the table on page 280 for the register bit definitions
Status System Commands Standard Event Register Commands
See the table on page 282 for the register bit definitions
Status System Commands Alarm Register Commands
STATusALARmCONDition?
STATusALARmEVENt?
STATusALARmENABle enablevalue STATusALARmENABle?
Status System Commands Standard Operation Register Commands
STATusOPERationENABle enablevalue STATusOPERationENABle?
See the table on page 285 for the register bit definitions
STATusOPERationCONDition?
Miscellaneous Status Register Commands
STATusPRESet
DATAPOINtsEVENtTHReshold numrdgs DATAPOINtsEVENtTHReshold?
PSC
Calibration Commands
CALibration?
CALibrationCOUNt?
CALibrationSECureCODE newcode
CALibrationSECureSTATe OFFON,code CALibrationSECureSTATe?
CALibrationSTRing quotedstring
CALibrationSTRing?
CALibrationVALue value CALibrationVALue?
Service-Related Commands
DIAGnosticDMMCYCLes?
DIAGnosticDMMCYCLesCLEar
DIAGnosticRELayCYCLes? @chlist
WAI
An Introduction to the Scpi Language
An Introduction to the Scpi Language
Command Format Used in This Manual
Command Separators
Using the MIN and MAX Parameters
Scpi Command Terminators
Querying Parameter Settings
IEEE-488.2 Common Commands
Scpi Parameter Types
301
Using Device Clear
Using Device Clear
Error Messages
Error Messages
Execution Errors
Execution Errors 114
121
123
124
Execution Errors 168
178
211
213
Execution Errors 230
310
350
410
Instrument Errors
225
Instrument Errors 221
226
Instrument Errors 261
271
272
281
305
Instrument Errors 303
306
Instrument Errors 501
502
511
512
Self-Test Errors
601
602
603
Calibration Errors
Calibration Errors 710
720
721
722
Plug-In Module Errors
901
902
903
318
Application Programs
Application Programs
Example Programs for Excel
Example Programs for Excel
Excel 7.0 Example takeReadings Macro
VISAaddr =
323
324
Excel 7.0 Example ScanChannels Macro
Dim columnIndex As Integer
326
327
Example Programs for C and C++
Example Programs for C and C++
++ Example dacout.c
++ Example statreg.c
331
332
Tutorial
Matrix Switching, starting on
Tutorial
System Cabling and Connections
Cable Specifications
System Cabling and Connections
Ft 2 conductors
Cable Type Nominal Impedance Capacitance Attenuation
Grounding Techniques
Separation of High-Level and Low-Level Signals
Shielding Techniques
Sources of System Cabling Errors
Copper-to Approx. ∝V / C
Where
Discussion of integration time
Voltage Measured = √ V + Noise
Measurement Fundamentals
Measurement Fundamentals
Integration Time PLCs
NMR
Temperature Measurements
RTD
Measurement Fundamentals
Internal DMM
Internal DMM Ice Bath
349
Measurement Fundamentals
Temperature Probe Type Pos + Lead Neg Lead
Iron Constantan
Nickel-Chromium Nickel-Aluminum
Copper Constantan
Sources of Error in Thermocouple Measurements
Without Shield
DC Voltage Measurements
Sources of Error in DC Voltage Measurements
Or greater see page 103 for a discussion of integration time
Connection a Connection B
Rs + Ri
358
AC Voltage Measurements
360
361
Sources of Error in AC Voltage Measurements
363
For low frequencies
+ Noise
Measurement Fundamentals
Current Measurements
Current measurements are allowed only on the 34901A module
Sources of Error in DC Current Measurements
Sources of Error in AC Current Measurements
Resistance Measurements
370
Measurement Fundamentals
Sources of Error in Resistance Measurements
Insulating Material Resistance Range Moisture Absorbing
Strain Gage Measurements
HI Source HI Sense LO Sense LO Source
Strain
DMM Sensitivity
Frequency and Period Measurements
Sources of Error in Frequency and Period Measurements
Low-Level Signal Multiplexing and Switching
Low-Level Signal Multiplexing and Switching
One-Wire Single-Ended Multiplexers
Two-Wire Multiplexers
Four-Wire Multiplexers
Signal Routing and Multiplexing
Sources of Error in Multiplexing and Switching
Module Bank
Actuators and General-Purpose Switching
Actuators and General-Purpose Switching
Snubber Circuits
RC Protection Networks
Using Varistors
Using Attenuators
Matrix Switching
Matrix Switching
Combining Matrices
Matrix Module
RF Signal Multiplexing
RF Signal Multiplexing
Sources of Error in RF Switching
Insertion Loss 75Ω
Digital Input
Multifunction Module
Multifunction Module
Digital Output
Using an External Pull-Up
Driving External Switches
Totalizer
Totalizer Errors
Voltage DAC Output
DAC Errors
Relay Life and Preventative Maintenance
Relay Life and Preventative Maintenance
Relay Life
Relay Load
Switching Frequency
Replacement Strategy
402
Specifications
DC, Resistance, and Temperature Accuracy Specifications
DC, Resistance, and Temperature Accuracy Specifications
DC Measurement and Operating Characteristics
DC Measurement and Operating Characteristics
AC Accuracy Specifications
AC Accuracy Specifications
AC Measurement and Operating Characteristics
AC Measurement and Operating Characteristics
Measurement Rates and System Characteristics
Measurement Rates and System Characteristics
Module Specifications
Module Specifications
34905A, 34906A
Typical AC Performance Graphs
Typical AC Performance Graphs
Insertion Loss 75 Ω
Crosstalk 75 Ω
Software Specifications
34907A
Product and Module Dimensions
Product and Module Dimensions
To Calculate Total Measurement Error
To Calculate Total Measurement Error
415
Interpreting Internal DMM Specifications
Interpreting Internal DMM Specifications
Number of Digits and Overrange
Sensitivity
Accuracy
Resolution
Criteria
Hour Accuracy
Temperature Coefficients
Day and 1-Year Accuracy
Configuring for Highest Accuracy Measurements
Configuring for Highest Accuracy Measurements
420
Totalize Threshold jumper, 135, 175 totalizer reset mode
Index
Index
423
424
425
426
427
428
429
430
431
432
433
434
U T I O N