Tyco 1847L Series manual Important Safety Instructions

Page 47

Attention, consult accompanying documents with respect to electric Shock, fire and mechanical hazard only in according with UL2601-1 and CAN/CSA C22.2 No. 601.1. medical equipment certified for Canada, E230760,IEC60601-1.

Manufacturing

Elo TouchSystems Inc.

6500 Kaiser Dr. Fremont, CA 94555 U.S.A.

Rating:

LCD Monitor Input Rating: 12Vdc, 4A

adapter Chi Sam Electronic Enterprise Co., Ltd. Type CH-1204M rating: I/P: 100-240Vac, 47-63Hz, Output: 12Vdc, 4A

Classification

1.Protection against electric shock (5.1, 5.2) Class I equipment, no applied parts

2.Protection against harmful ingress of water (5.3) Ordinary Equipment (IPXO)

3.Degree of safety in the presence of a flammable anesthetics mixture with air or with oxygen or with nitrous oxide (5.5)

Not suitable for use in the presence of a flammable anesthetics mixture with air or with oxygen or with nitrous oxide.

4.Model of operation (5.6) Continuous

Explanation Symbol

1. : DC Current

2. : ON/ OFF switch

Important Safety Instructions

Model: ET184XL-XXXB*, ET186XL-XXXB*.

1.Read the Safety Instruction carefully and keep it for use later.

2.Be aware of all warning and instruction signs marked on the products.

3.When cleaning, cut off the electrical supply at all times, Never use liquid or Aerosol detergent, use a damp rag instead.

4.The LCD Monitor shall be operated according the following specification.

5.The adapter can be operated at an ambient temperature 10~40 C,RH 10%~90% maximum.

6.The adapter shall be operated in dry locations to protect from moisture.

7.Keep this product stable at any time. Otherwise the product may fall, hazarding the safety and causing serious damage.

8.Do not clog apertures on the product's coating required for ventilation purpose.

Do not install this product in built_in installation without proper ventilation system. Always keep this product away from all kinds of heat sources.

9.The power source used for this product must match one marked on the product's label. Please consult your dealer or local technician if you have any doubt.

10.In order to avoid the electric shock hazard in use, a 3-wire type plug with a grounding pin is provided. Do not use any kind of plug without grounding..

11.Do not lay the power cord on the pathway or rest anything heavy on it

12.Check the current flow tolerance of an extension cord before use, make sure that the total amount of the product's current flow does not exceed the limit. A current flow of a wall outlet can never exceed up to 15 amperes.

13.Do not insert objects or pour liquid into this product through it's casting apertures. It may touch a high voltage area causing electric shock or short the circuit.

14.Compliance with IEC EN 60601-1-2 and UL 2601-1, CLASS I

15.Transportation temperature/humidity: -20 C ~+60 C /RH 10% ~90%

16.The LCD Monitor Model ET184XL-XXXB*, ET186XL-XXXB*is intended use for medical purpose.

17.For repair of equipment parts designated by the manufacturer as repairable provided.

42 E l o E n t u i t i v e T o u c h m o n i t o r U s e r G u i d e

Image 47
Contents Elo Entuitive Touchmonitor User Guide Elo Entuitive Touchmonitor User Guide Copyright 2003 Elo TouchSystems Inc. All Rights Reserved Page Table of Contents About the Product PrecautionsPage Unpacking Your Touchmonitor Installation and SetupProduct Overview Attaching the L-Brackets TOP Panel Cut Outs1847L Option 1847L Option 1847L with 819602-000 Option Top Serial or USB Connection Touch Interface ConnectionConnecting the Video Cable Connecting the Serial or USB Touchscreen Cable Connecting the OSD Remote Cable Connecting the Power Cable Vesa Mount on Your Touchmonitor Optimizing the LCD DisplayVesa Mounting Interface Installing the Driver Software Installing the Serial Touch Driver Installing the Serial Touch Driver for MS-DOS and Windows Page Page Using the On-Screen Display OSD Menus About Touchmonitor AdjustmentsControl Function Remote OSD ButtonsAuto Contrast OSD Menu FunctionSaturation HUE Flesh Tone Solutions to Common Problems TroubleshootingPage Native Resolution Page Touchmonitor Safety Care and Handling of Your Touchmonitor Compatible Video Modes Technical SpecificationsTouchmonitor Specifications Optical Electrostatic Positional Accuracy Side View Regulatory Information Page CE230760 US Important Safety Instructions Warranty Page Index D e x 4 Check out Elos Web site