Omega Engineering DRN-PS-750 manual Safety Considerations, Unpacking & Inspection

Page 16
Note

SAFETY CONSIDERATIONS

This device is marked with the international Caution symbol. It is important to read this manual before installing or commissioning this device as it contains important information relating to Safety and EMC (Electromagnetic Compatibility).

Unpacking & Inspection

Unpack the instrument and inspect for obvious shipping damage. Do not attempt to operate the unit if damage is found.

This instrument is a panel mount device protected in accordance with Class I of EN 61010 (115/230 AC power connections). Installation of this instrument should be done by Qualified personnel. In order to ensure safe operation, the following instructions should be followed.

This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947Ð1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the mains supply cord.

Furthermore, to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment, an overcurrent protection device shall be installed.

¥The Protective Conductor must be connected for safety reasons. Check that the power

cable has the proper Earth wire, and it is properly connected. It is not safe to operate this unit without the Protective Conductor Terminal connected.

¥Do not exceed voltage rating on the label located on the top of the instrument housing.

Note ¥ Always disconnect power before changing signal and power connections.

¥Do not use this instrument on a work bench without its case for safety reasons.

¥Do not operate this instrument in flammable or explosive atmospheres.

¥Do not expose this instrument to rain or moisture.

EMC Considerations

¥Whenever EMC is an issue, always use shielded cables.

¥Never run signal and power wires in the same conduit.

¥Use signal wire connections with twisted-pair cables.

¥Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.

14

DNR-PS Ñ DIN Rail Power Supplies

Image 16
Contents Page USA Contents General Information and Features FeaturesApplicable Standards Input / Output IsolationInstallation and Removal To change voltage jumpers Ñ Main Board Power Jumpers Installation Removal Input and Output Connections Specifications Temperature RangesInput Range Temperature Derating Derating Graph / Block Diagram Dimensions Drawing Ñ DimensionsSafety Considerations Unpacking & InspectionReturn Requests / Inquiries OMEGAÉOf Course