COMMAND REFERENCE

LDX-3232 Device-Dependent Command Reference

C H A P T E R 4

 

Circuit condition is fleeting and may be missed via the LAS:COND? query. Therefore, the

 

user should test for the Open Circuit Event via the LAS:EVEnt? query.

 

The LASER condition status is constantly changing, while the event status is only

 

cleared when the event status is read or the *CLS command is issued.

Examples

"LAS:COND?" -response: 513, means that the LASER limit current and out of tolerance

 

LASER conditions currently exist.

 

"Radix Hex; Laser:Cond?" -response: #H108, means that the LASER Output shorted

 

and Power limit conditions currently exist.

LASer:DEC

FRONT PANEL

 

 

REMOTE

The LASer:DEC command decrements the selected laser control mode setpoint by one or more steps. Optional parameters allow multiple steps to be decremented and the time (in milliseconds) between decrements to be set, respectively. This is an overlapped command. Other commands may be executed while this function is in progress.

Syntax Diagram

:

LAS

:

DEC

<white

space>

<nrf

value>

,

<nrf

value>

Parameters

0, 1, or 2. The first optional parameter is the number of steps, and the second optional

 

parameter is the number of milliseconds between steps.

Notes

The decremental default amount is one step. The step size can be edited via the

 

LAS:STEP command. LDX-3232 default values are 0.1 mA, 0.1 mW, or 1 µA

 

(if CALMD = 0).

 

If the first optional parameter is used, but not the second, the user may decrement the

 

LASER setpoint by a multiple of the LAS:STEP size, without changing the LAS:STEP

 

size.

 

If the both optional parameters are used, the user may create an automated stepping

 

ramp function for the LASER output.

 

If the first optional parameter is entered as zero, "LAS:DEC 0", the command will do

 

nothing.

 

The minimum time to complete one decrement is about 20 ms. Therefore, values for the

 

second optional parameter (time between decrements) have a practical minimum of 20.

Examples

"LAS:MODE:I; LAS:STEP 3; LAS:DEC" -action: The laser source current setpoint is

 

decremented by 0.3 mA.

 

"LAS:MODE:I; LAS:STEP 3; LAS:DEC 3" -action: The laser source current setpoint is

 

decremented by three times the amount described in the first example.

 

"LAS:MODE:I; LAS:STEP 3; LAS:DEC 3,5000" -action: The laser source current

 

setpoint is decremented by the amount described in the first example, three times, with

 

five seconds between decremental steps.

04_06

LDX-3232

47

Image 63
Contents User’s Guide Page Table of Contents Remote Operation Command Reference Iv LDX-3232 List of Figures Vi LDX-3232 List of Tables Viii LDX-3232 General Safety Considerations Safety Information and the ManualSafety Marking Symbols Safety SymbolsWarranty Claims for Shipping Damage Comments, Suggestions, and Problems Description of the problemModel Number Serial Number End-User Name Company Phone Fax Xiv LDX-3232 Product Overview Introduction and SpecificationsGrounding Requirements Installing Your LDX-3232 High Compliance Current SourceAC Line Power Requirements Initial InspectionOperating the LDX-3232 Precision Current Source Gpib ConnectorTilt-Foot Adjustment Gpib AddressLDX-3232 Front View Maintenance Specifications Type Pulse Width Delay TTL13 µs 12 ms Description Model Number Available Options and AccessoriesPower On Sequence Applying Power to Your LDX-3232Power On State Connecting to the LaserCommon Laser Cathode Photodiode Anode Interlock Connections Back Panel LD ConnectorSetting the PD Bias Photodiode ConnectionsGrounding Considerations Operating a Laser in Constant Current I Mode Front Panel OperationDisplay Setting Up the Precision Current Source E R a T I O N Entering Photodiode Responsivity Values Operating a Laser in Constant Power P ModeMethod to calculate the responsivity is suggested below LDX-3232 Error Indicators Error IndicatorsError Indicators LDX-3232 Changing the Gpib Address Reading the Gpib AddressLDX-3232 Current Source Command Set Command SyntaxChanging Operation from Local to Remote OFF OLD NEW True False SET Reset LASLDI? LDX-3232 Command Path Structure Command PathsIEEE488.2 Command Commands LDX-3232 Frequently Used Commands NoneLIM Status Reporting Event and Condition RegistersOperation Complete Definition LDX-3232 Status Reporting Schematic Diagram Output Off Register Disabled 10 N/A Enabled 11 N/A 12 N/A 13 N/A 14 N/A 15 N/ALASerENABleOUTOFF nrf LASerENABleOUTOFF? Error Messages Command Timing and CompletionResponse Message Error Code ClassificationsMode Used to enter the laser voltage calibration mode Used to create a delay in the execution of furtherCommands Used to enter the laser current calibration modeReturns the value of the LAS conditions enable register Returns status of photodiode monitor power displayEnables Laser display to show parameter values Sets the enable register for Laser conditionsOr P optical power Used to return the optical power from monitorUsed to set the constant optical power setpoint, if Used to enable/disable the front panel modulation inputUsed to return the Laser setpoint tolerance value Returns the status of the Term commandPowered up Returns the elapsed time since the timer was last resetCommand Description Format LDX-3232 Device-Dependent CommandsDelay LDX-3232 Device-Dependent Command ReferenceERRors? LASerCALMDCalmd LAS Calpd LASerCALMD? Responsivity CAL PD parameter setting Syntax DiagramSelected. This procedure is outlined in Chapter LASerCALLDICalibration for that bandwidth LAS CAL LDV LASerCALLDI?LASerCALLDV LDI LAS CALTrue measured value. This procedure is outlined in Chapter LASerCALLDV?LASerCALMDI LAS CAL IPD LASerCALMDI?LASerCOND? MDILASerDEC LASDIS? LASerDISplayLASerDISplay? LAS DISLDI LASerDISplayLDILASerDISplayLDI? LasdisldiLASDISLDV? Voltage measurement Syntax DiagramLASerDISplayLDV LASerDISplayLDV?LAS DIS IPD LASerDISplayMDILASerDISplayMDI? LAS DIS MDILAS DIS MDP LASerDISplayMDPLASerDISplayMDP? PPDLAS DIS SET LASerDISplayPARAMLASerDISplaySET LAS DIS ParamLASerENABleCOND LASerDISplaySET?LASDISPLAYSET? LASerENABleCOND? BitLAS Enab Cond Lasenabevent LASerENABleEVEntAn nrf value whose sum represents the bits that are enabled Action enables the Laser status event register so thatLAS Enab EVE LASerENABleEVEnt?LASerENABleOUTOFF LAS Enab Outoff LASerENABleOUTOFF?LAS EVE LASerENABleEVEnt commandCommand while the condition status is constantly changing LASerEVEnt?Function for the Laser output LASerINCLAS LDI This measurement is updated approximately once every 600 msLASerLDI LASerLDI?LAS LDV LASerLDVLASerLDV? For LDV calibrationLAS LIM LASerLIMitI2LASerLIMitI2? Limit in use depends on the Laser output range selectionLASLIMI4? Displayed, and then releasing the SET switchLASerLIMitI4 LASerLIMitI4?LASerLIMitMDP LASerLIMitMDP?MDP LAS LIM LASerLIMitV? LASerLIMitVMDP P PD setpoint would be used LASerMDILASerMDI? MDI LAS IPDLASerMDLN LASerMDLN?LAS Mdln LASerMDP LASerMDP?MDP LAS LASerMODEIHBW LAS ModeLAS Mode Ihbw LASerMODE?Lasmodeilbw LAS Mode MDPLASerMODEILBW LASerMODEMDPLASerOUTput LASerOUTput?LAS OUT LASRAN? LASerRANgeLASerRANge? LAS RANLASerSETMDI? Where the response is the constant I setpoint value, in mA30 µA, for use in constant P mode with Calpd = LASerSETLDI?100 µA Where the response is the constand PPD setpoint, in mWLASerSETMDP? LASerSTEPLAS TOL LASerTOLerance commandLASerSTEP? LASerTOLeranceMES LASerTOLerance?MESsage Second is the time window, in secondsRADix MESsage?Means hexadecimal, and OCT means octal Term 0 command is sent or the LDX-3232 is powered offRADix? RADLDX-3232 was last powered up Time clock is independent of the Timer clockTime Seconds since the last TIMER? query was issued Since the last TIMER? query was issuedTimer Last TIMER? query is returnedSaving and Recalling from the Front Panel Functions and FeaturesSaving and Recalling Under Remote Operation Using the Laser Compliance Voltage AdjustmentLDX-3232 Rear Panel Using the LDX-3232 Current Source’s Trigger FunctionBNC Interlock Usage Example Modulating the Laser Current SourceModulating the Laser Current Source Functions and Features Calibration Overview Calibration and TroubleshootingDescription Manufacturer/Model Specification Recommended EquipmentDMM Current Limit Local Calibration of the LDX-3232 Current SourceCurrent Source Calibration RangeIPD Current Calibration IPD Setpoint Constant powerCAL PD Set equal to zeroSet to FS full scale of range Laser ModeLaser Forward Voltage Measurement Calibration 2000 mAGpib Command Remote Calibration of the LDX-3232 Current SourceLimit Lasldi nrf value command IPD Setpoint Function State Gpib Command Current source must be recalibrated Troubleshooting GuideGeneral EEPROM, Lllll will be displayed on the display the laserCommand Feature can be defeated via the Lasenableoutoff101 Response Message