A.Pycnometer Not Getting Vacuum
1.Check 'O' ring to make sure it is clean and in good shape. If not, contact ELE/Soiltest for a replacement 'O' ring.
2.Check to make sure the water inlet valve is on tight.
3.Tighten the stopcock nut (at bottom of stopcock) of the water inlet valve.
B.Air Bubbles Entering from the Water Inlet Valve
1.Check to make sure the water inlet valve is on tight.
2.Tighten the stopcock nut (at bottom of stopcock) of the water inlet valve.
C.Clogged
D.Too Many Air Bubbles in Water
1.Air bubbles come from dissolved air in water. Check for this by opening water inlet valve (keeping the water inlet hose connected to the water supply) after you have gone through Step 8 of the operating procedure. The air bubbles should dissolve back in water at atmospheric pressure.
2.Try filling the pycnometer with boiled water (after it has been cooled).
If problem still persists, call ELE/Soiltest Service Department for assistance.