80TK Calibration Procedure
A calibration cycle of 1 year is recommended to maintain the 80TK within specifications. The required equipment is listed below. Use Calibration Procedure 1 to calibrate the 80TK for use with any
Location of Calibration Adjustments
Calibration Procedure 1.
Use this procedure to calibrate the 80TK:
1.Verify the condition of the battery as described above and replace the battery if necessary. Follow the Battery Replacement procedure to disassemble the 80TK, but do not remove the battery or reassemble the 80TK.
2.Allow 80TK and the
3.Use the shorting bar to short the 80TK thermocouple input terminals.
4.Plug the 80TK into the dc volts input of the multimeter. Set the multimeter to the 300 mV or equivalent range.
5.Set the switch on the 80TK to CELSIUS position. Adjust R4 to half rotation, then adjust R5 until the reading on the
multimeter matches the reading on the mercury reference thermometer ± 5.0°C. Now adjust R4 until the two readings are within ± 0.1°C.
Note
If the mercury reference thermometer displays °F, use the following conversion to get °C:
[5/9 (temp in
6.Set the switch on the 80TK to the FAHRENHEIT (center) position.
7.Adjust R16 until the reading on the multimeter matches the reading on the mercury reference thermometer ± 0.2°F.
Note
If the mercury reference thermometer displays °C, use the following conversion to get °F:[9/5(temp in °C) + 32] = (temp in °F)
8.Set 80TK switch to the OFF position. Remove 80TK from multimeter and remove any connections to the 80TK thermocouple input terminals.
9.Reassemble the 80TK.
10.The 80TK is now calibrated.
Calibration Procedure 2.
Use this procedure if the 80TK will be used exclusively with one
Perform Calibration Procedure 1. Substitute the following for step 3 in Calibration Procedure 1:
3.Connect the thermocouple to the 80TK thermocouple input, and place the thermocouple into the room temperature bath along with the mercury reference thermometer.
Required Equipment
Refer to Table 1 for equipment required to calibrate the 80TK.
Instrument
Mercury Reference Thermometer
Dewar Flask and Cap(water bath)
Digital Voltmeter
Shorting bar
OR
Table 1. Required Equipment
Minimum Specification
0.1°C resolution
2 pint capacity, filled with water
300 mV range, 100 μV resolution
<1 ohm
meets NBS standards(with subminiature K plug)
Recommended Model
Princo Model
Fluke Model 87 Series III
Wire
Fluke
To contact Fluke, call one of the following telephone numbers:
USA: | Japan: | |
Canada: | Singapore: | |
Europe: | +31 | Anywhere in the world: |
Or, visit Fluke’s Web site at www.fluke.com. |
| |
| Fluke Corporation | Fluke Europe B.V. |
| P.O. Box 9090 | P.O. Box 1186 |
| Everett WA | 5602 B.D. Eindhoven |
| USA | The Netherlands |