3M 1292 manual Product Description, Indications, Contraindications, Precautions

Page 1

3M™ Attest™ 1292 Rapid Readout Biological Indicator

Product Description

The 3M™ Attest™ 1292 Rapid Readout Biological Indicator (brown cap) is a dual readout biological indicator system specifically designed for rapid and reliable monitoring of steam sterilization processes when used in conjunction with the 3M™ Attest™ 290 Autoreader or the 3M™ Attest™ Autoreader 390.

The Attest™ 1292 Rapid Readout Biological Indicator detects the presence of Geobacillus stearothermophilus by detecting the activity of alphaglucosidase, an enzyme present within the organism. The presence of the enzyme is detected by reading fluorescence produced by the enzymatic breakdown of a nonfluorescent substrate. This creates a fluorescence change, which is detected by the autoreader. A fluorescence change indicates a steam sterilization process failure.

The Attest™ 1292 Rapid Readout Biological Indicator also detects the presence of G. stearothermophilus organisms by a visual color change reaction. Biochemical activity of the G. stearothermophilus organism produces acid byproducts that cause the media to change color from purple to yellow. A visual pH color change also indicates a steam sterilization process failure. Due to the high sensitivity of the 3hour fluorescent results, however, there is no advantage to incubating the Attest™ 1292 Rapid Readout Biological Indicator beyond 3 hours.

Indications

Use the Attest™ 1292 Rapid Readout Biological Indicator to monitor:

1.250°F (121°C) gravity steam sterilization cycles.

2.270°F (132°C) vacuum assisted steam sterilization cycles.

Contraindications

None.

Warning

There is a glass ampoule inside the plastic vial of the biological indicator. To avoid the risk of serious injury from flying debris due to a ruptured biological indicator:

Allow the biological indicator to cool for the recommended time period before crushing. Crushing or excessive handling of the biological indicator before cooling may cause the glass ampoule to burst.

Wear safety glasses and gloves when removing the biological indicator from the sterilizer.

Wear safety glasses when crushing the biological indicator.

Handle the biological indicator by the cap when crushing and tapping.

Do not use your fingers to crush the glass ampoule.

Do not roll the biological indicator between fingers to wet the spore strip.

Precautions

Do not use the Attest™ 1292 Rapid Readout Biological Indicator to monitor:

1.250°F (121°C) vacuum assisted steam sterilization cycles.

2.270°F (132°C) gravity steam sterilization cycles.

3.Dry heat, chemical vapor, ethylene oxide or other low temperature sterilization processes.

Monitoring Frequency

Follow facility Policies and Procedures which should specify a biological indicator monitoring frequency compliant with professional association recommended practices and/or national guidelines and standards. As a best practice and to provide optimal patient safety, 3M recommends that every steam sterilization load be monitored with an appropriate biological indicator.

Directions for Use

1.Identify the Attest™ 1292 biological indicator by writing the sterilizer and load ID number and processing date on the indicator label. Do not place another label or indicator tape on the vial or on the cap.

2.Place the Attest™ 1292 biological indicator in an appropriate process challenge device according to recommended practices. Do not place the Attest™ 1292 biological indicator in direct contact with a chemical indicator. Fluorescent residue could transfer to the biological indicator and affect the result.

Appropriate process challenge devices for loads containing:

a.Fabric packs run at 250°F (121°C) for ≥40 minutes in a gravity displacement cycle or at 270°F (132°) for ≥4 minutes in a vacuum assisted cycle:

-Place an Attest™ 1292 biological indicator in an AAMI 16 towel pack.

b.Wrapped hard goods run at 250°F (121°C) for ≥20 minutes in a gravity displacement cycle:

-Place an Attest™ 1292 biological indicator in a wrapped hard good item (e.g., instrument set) from the load.

c.Wrapped hard goods run at 270°F (132°C) for ≥4 minutes in a vacuum assisted sterilizer:

-Place an Attest™ 1292 biological indicator in a wrapped hard good item (e.g., instrument set) from the load.

-Place an Attest™ 1292 biological indicator in an AAMI 16 towel pack.

d.Single wrapped hard goods run at 270°F (132°C) for ≥4 minutes in an express vacuum assisted cycle:

-Place an Attest™ 1292 biological indicator in a single wrapped hard goods item (e.g., instrument tray) from the load. Check recommendations of the manufacturer before processing items with lumens or porous materials.

e.Unwrapped hard goods with no porous items run at 250°F (121°C) in a gravity displacement cycle for ≥ 15 minutes or 270°F (132°C) for ≥ 3 minutes in a vacuum

assisted sterilizer:

-Place an Attest™ 1292 biological indicator in unwrapped instrument hard good item (e.g., instrument set) from the load. AAMI suggests placing a biological indicator in an empty tray.

f.Unwrapped metal hard goods with porous items run at 270°F (132°C) for ≥ 4 minutes in a vacuum assisted sterilizer:

-Place an Attest™ 1292 biological indicator in an unwrapped instrument hard goods item (e.g., instrument tray) from the load, including porous items. AAMI suggests placing a biological indicator in an empty tray but include porous items if applicable.

g.Container systems run at 250°F (121°C) for ≥ 40 minutes in a gravity cycle or 270°F (132°C) for ≥ 4 minutes in a vacuum assisted cycle:

-Place the Attest™ 1292 biological indicator in the areas determined by product testing to provide the greatest challenge to the sterilization process.

3.Place the test tray or package in the most challenging area for the sterilant. This is typically on the bottom shelf, near the door and over the drain.

4.Process the load according to recommended practices.

5.After completion of the cycle, fully open the sterilizer door for a minimum of 5 minutes prior to removing the Attest™ 1292 biological indicator.

6.When the biological indicator is not contained in a test pack or other heat absorbing packaging material, remove the biological indicator from the sterilizer and allow it to cool for an additional 10 minutes prior to crushing.

Image 1
Contents Product Description PrecautionsIndications ContraindicationsStorage/ Shelf Life DisposalPH Color Change Result Optional Growth Fluorescence Explanation of SymbolsMade in U.S.A. by 3M Health Care Conway Ave

1292 specifications

The 3M 1292 is a high-performance adhesive transfer tape that has gained considerable recognition in various industrial and commercial applications. Known for its reliability and versatility, the 3M 1292 is designed to meet the demands of different industries while providing exceptional bonding strength.

One of the main features of the 3M 1292 is its impressive adhesive formulation. This transfer tape utilizes an aggressive acrylic adhesive that provides excellent initial tack and long-term holding power. The adhesive is designed to bond a wide variety of substrates, including metals, plastics, and composites, making it suitable for diverse applications. Its ability to adhere to both smooth and rough surfaces allows for greater flexibility in design and manufacturing processes.

The advanced technologies employed in the production of the 3M 1292 enhance its performance. The tape features a thin transfer film that ensures a clean and precise application. This thin profile minimizes bulk while maintaining strength, allowing for a seamless finish in applications such as signage, displays, and nameplates. Additionally, the transfer tape is engineered to resist temperature fluctuations and various environmental factors, ensuring the integrity of the bond even in challenging conditions.

Another important characteristic of the 3M 1292 is its ease of use. It comes in convenient rolls that can be easily cut to size, enabling users to apply the tape according to their specific needs. The tape's easy-release liner simplifies the application process, allowing for efficient handling during manufacturing or assembly. This user-friendly design contributes to reduced labor costs and enhances productivity in various workflows.

The 3M 1292 is also notable for its durability. The tape is resistant to UV light, moisture, and chemicals, making it suitable for outdoor applications without compromising its adhesion properties. This durability extends the lifespan of bonded products and reduces the need for frequent maintenance or replacement.

In summary, the 3M 1292 adhesive transfer tape stands out due to its aggressive adhesive formulation, advanced production technologies, user-friendly application, and remarkable durability. These features make it a top choice for industries ranging from signage and automotive to electronics and building materials, effectively supporting a multitude of bonding requirements.