Series 900 Single & Dual Instruction Manual
The VALVE PURGE function is activated by the VALVE TP (test point) pin. When this pin is connected to common, the valve is fully open regardless of setpoint. This pin must source 150 mA of current when driven to common. If a voltmeter is connected between this pin and
The
CAUTION! If a relay is being driven by an alarm, A protection diode must be connected across the relay coil with reversed polarity.
A high and low alarm may be connected in parallel to make a window alarm. Whenever the flow is inside of the window defined by the two alarms, the alarm is inactive (OFF). If the flow is outside of the window, the alarm is active (ON). See Appendix D for suggested alarm configurations.
There are two plastic programming jumpers plugged into a header on the back panel. For each channel they determine whether the
4.3.3 High/Low Flow Alarm Outputs
4.4
Automatic/Manual Setpoint Selection (Controllers)
Page 6