Graham Field 765000 user manual Pump Front Panel, Front panel, first row

Page 9

PUMP: FRONT PANEL

Front panel of pump is shown below. Descriptions of buttons and indicators follow.

Auto-Adjust button;

Low pressure

Tuning button;

indicator

 

indicators at right of button

 

indicators at left of button

 

 

Therapy adjust button;

 

 

Maxfirm button

indicators at right of button

 

 

 

Panel lock button;

 

 

Service indicator

 

 

 

indicator at right of button

 

Power failure indicator

Power (ON/OFF) button;

Alarm mute button

indicator at right of button

 

 

 

Front panel, first row

Auto-Adjust button: The auto-adjust button adjusts the mattress pressure automatically based on the patient's weight. To remind the user that the auto- adjust process is ongoing, the auto-adjust indicator LEDs will periodically flash from left to right until the Ready LED illuminates to show the process is complete. Three conditions trigger the auto-adjust function:

1.Press the auto-adjust button for two seconds; the mattress system automatically readjusts pressure setting.

2. When initial mattress inflation is complete, the mattress

system automatically starts the auto-adjust function.

3.When the mattress system detects a significant change in patient's weight on the mattress (e.g. patient egress or ingress), the system automatically starts the auto- adjust function.

When auto-adjust is complete, the Ready indicator will illuminate. To disable the auto-adjust function, press the auto-adjust button until auto-adjust indicators de- illuminate.

Info: During the auto-adjust operation, it is normal for the mattress system to cycle through a series of inflation and deflation.

765000-INS-LAB-RevC11

9

Image 9
Contents 765000 Contents Introduction Intended USE of this DeviceContraindication To reduce the risk of electrocution Important Safety PrecautionsINS-LAB-RevC11 INS-LAB-RevC11 Equipment Symbols Pump and Mattress System Product DescriptionPump Front Pump RearFront panel, first row Pump Front PanelFront panel, second row Front panel, third row INS-LAB-RevC11 Installation Pump and Mattress SetupUnpacking INS-LAB-RevC11 Operation Getting StartedQuick Disconnect CPR Function Transport CAPMaintenance CleaningGeneral Maintenance Fuse replacement Air filter replacementStorage Pump Specification SpecificationsLimited Warranty Appendix a EMC Information INS-LAB-RevC11 =1.2 P = 2.3 P GF Health Products, Inc