Transporting & Storing VIAspire Portables
Preparing for Shipping
Prior to shipping the VIAspire Portable 300 or 600, remove the batteries and empty any residual oxygen from the unit. Allow the portable to sit for at least 8 hours to allow it to empty.
Storage
Store the VIAspire Portable 300 or 600 in a cool, clean, dry area when not in use. Avoid storing in direct sunlight.
Battery disposal
Dispose of all alkaline batteries in accordance with local, state, and federal regulations.
Specifications
Batteries | 4 |
|
Required | (alkaline only) |
|
|
|
|
Oxygen | 22 – 25 psig |
|
Pressure |
|
|
|
|
|
LOX Content | Model 300 | 0.33 liters (approx) |
Volume | Model 600 | 0.6 liters (approx) |
| ||
|
|
|
Weight | Model 300 | 3.5 lb |
(empty) | Model 600 | 4.5 lb |
| ||
|
|
|
Temperature | Storage/Transport | |
| Operating | + 5 to 40°C |
|
|
|
Humidity | Storage | 0 to 95% RH, |
Operating | 15 to 95% RH, | |
| ||
|
|
|
18 VIAspire Oxygen Portable Manual |