Iron (0.10 - 5.00 mg/L Fe) | HCT 159 |
UniCell™ Test
Scope and Application: For drinking water, raw water,
US/GB
SUMMARY OF METHOD
Iron(II) ions form an
SHELF LIFE OF TEST REAGENT
The test reagents are stable at room temperature for 30 months as indicated by the expiration date on the package.
SAMPLE COLLECTION, PRESERVATION AND STORAGE
Collect samples in
Before analysis, adjust the pH to between 3 and 5 with
5.0 N sodium hydroxide standard solution. Do not exceed pH 5
as iron may precipitate. Correct the test results for volume additions.
INTERFERENCES
The ions listed in the table have been individually checked up to the given concentrations. Cumulative effects and the influence of other ions have not been evaluated. There is no interference from:
METHOD PERFORMANCE
Precision
In a single laboratory, using a standard solution of 3.00 mg/L Fe and two representative lots of reagent, a single operator obtained
a standard deviation of +/- 0.15 mg/L on a total of 14 samples.
TIPS AND TECHNIQUES
1. Adjust pH of preserved samples to between pH 3 – 5
before analysis. Water samples with pH < 2 affects the reagent. 2. The outside of the vial must be clean before placing it in the
instrument. Wipe it down with a damp towel, followed by a dry one.
3. For a proof of accuracy, use the standard solution or standard addition method. (See Accuracy Check).
4. See your Instrument’s Procedure Manual for instructions on adjusting the calibration curve if needed.
5. The temperature of the water sample and the sample vial should be between 15 – 25°C (59 – 77F).
SAFETY
Good safety habits and laboratory techniques should be used throughout the procedure. Consult the Material Safety Data Sheet for information specific to the reagents used.
1000 mg/L: Cl -
500 mg/L: Ca2+
100 mg/L: Ag+
70 mg/L: Cd 2+
50 mg/L: Co2+, Zn2+, Pb2+,
25 mg/L: Ni 2+
10 mg/L: Cu2+
5 mg/L: Sn2+
Higher amounts of copper, nickel, and tin cause
Total iron including undissolved iron and complexed iron can only be determined after digesting with the Metal Prep Set HCT 200. (Total iron measuring range 0.12 – 6.00 mg/L).
ACCURACY CHECK
Standard Solution Method
Prepare a 3.00 mg/L Fe standard solution by pipetting 0.30 ml of 1000 mg/L Fe into a 100 ml volumetric flask. Dilute to the
mark with deionized water. Stopper and invert to mix. Prepare this solution daily. Perform the iron procedure as described.
Standard Addition Method
(a)Measure 100 ml of sample into three mixing cylinders.
(b)Using a pipette add 0.10, 0.20 and 0.30 ml of 1000 mg/L Fe standard solution to the three samples. Mix well.
(c)Analyze each sample using the procedure.
The Fe concentration should increase 1.00 mg/L for each 0.10 ml of standard added.
HCPE 159 B / Druckfarbe schwarz / 1