Omega Engineering DPG10L manual To Set Alarms

Page 7

M3367/0606

DPG10L

 

 

2.3To disengage Hold press and hold the “MODE” key until the reading starts flashing. Release the “MODE” key then press the “MODE” key once more, the display will read “HOLD”. Press the “ENTR” key, the display will read “NO H” Press the “ENTR” key once more and the display will return to reading the system pressure.

3.0To Set Alarms

3.1Press and hold the “MODE” key until the reading starts flashing, then Press the “MODE” key twice more. The display will read “AL 1”. Press the “ENTR” key to display the alarm setting. Pressing the “ENTR” key again will set the alarm and return the display to the normal mode. If the alarm setting is to be changed, the “up” and “down” keys can be used to increase or decrease the setting one count at a time, Pressing the “MODE” key and holding it down while Pressing “up” or “down” momentarily will change 10 counts for each key operation. If the “MODE” key and “up” or “down” are held down simultaneously, the setting will ‘FREE RUN’ until the key is released.

NOTE: While setting the alarm, only one segment of the analog scale is displayed and it corresponds to the % full scale of the alarm setting.

3.2When the desired alarm setting is displayed, Press “ENTR” to set and return to normal mode.

3.3To set alarm 2: Press and hold the “MODE” key until the reading starts flashing, then Press the “MODE” key three more times. The display will read “AL 2”. Press “ENTR” key to display the alarm setting. Pressing the “ENTR” key again will set the alarm and return the display to the normal mode. If the alarm setting is to be changed, the “up” and “down” keys can be used to increase or decrease the setting one count at a time, Pressing the “MODE” key and holding it down while Pressing “up” or “down” momentarily will change 10 counts for each key operation. If the “MODE” key and “up” or “down” are held down simultaneously, the setting will ‘FREE RUN’ until the key is released.

NOTE: While setting the alarm, only one segment of the analog scale is displayed and it corresponds to the % full scale of the alarm setting.

3.4When the desired alarm setting is displayed, Press “ENTR” to set and return to normal mode.



Image 7
Contents DPG10L PH/CONDUCTIVITY TemperatureTable of Contents Analog Output Signal Option Alarms OptionWarranty/Disclaimer Non-Warranty Repairs Warranty ReturnsReturn Requests/Inquiries To Zero or Tare OUT an existing offset InstructionsTo select Peak HI or Peak LO hold functions To Set Alarms Zero Trimming Battery ReplacementRe-calibration Alarms Option Analog Output Signal OptionDPG10L Output Options Typical 4-20 mA Circuit with Open Collector Alarms WiringThank You for Purchasing Omega Products Info@omega.com OMEGAnet On-Line Service Internet e-mail