DPG5000 | M3366/0606 |
|
|
• Instructions
NOTE: All units are factory calibrated prior to shipment.
1. Zero Trimming
NOTE: Some units are equipped with zero “knob”.
If it becomes necessary to
NOTE: A jewelers screwdriver or an eyeglass repair screwdriver (supplied) is a suitable instrument. Do not make changes to the Span adjustment (the “S” pot to the right of the push button) as part of the zero trimming. The Span should only be changed as part of the
2. Use of Auto-Tare Feature
When the DPG5000 is first turned on you will notice that the display flashes for the first 4 seconds. If the On/Off button is pushed and held during this period, any existing reading will be
3. Battery Replacement (9 Volt Type)
The battery can be replaced simply by removing the single screw at the top of the battery door. Remove the old battery, unplug the cable connector and replace with Eveready type 216 or equivalent. Replace the battery door and secure with the
NOTE: For best accuracy,