ADVANCED FEATURES

USING DIAGNOSTIC

Reset. With the many User Options available to customize the 250DX, it is easy to configure the unit in such a way that it is detrimental to a particular use. By using the Reset function, all variable or user- controlled features (except Filter) of the 250DX are returned to the factory settings.

The Reset function is an important first step in troubleshooting problems to ensure that a user selected setting is not the cause for the perceived problem.

USING THE DIAGNOSTIC FEATURE

The 250DX contains a powerful diagnostic tool that can aid in determining the cause of a problem.

To use the Diagnostic feature, select the Diagnostic feature on the start-up menu on the initial screen. Diagnostic can also be accessed through the Options menu.

Self Test. Initially, the 250DX will perform a self-test. This test confirms the operation of all internal circuitry. At the conclusion of the test, one of two messages will appear. PASSED indicates the internal test discovered no failures, FAILED indicates a significant internal problem was discovered and the unit requires factory service. A code number appears that tells repair personnel where the problem is.

Transducer Connection. The first connection tested is Transducer input. If a transducer is detected, the message CONNECTED appears. If a series of dashed lines is shown, no transducer is connected or there is

a problem with the transducer or cable.

If you are using multiple transducers, switch between transducers to confirm the connections for each.

24

Image 26
Contents Operations Manual Thank YOU Table of Contents Installation Overview Installation PreparationAccessories Parts SuppliedInstallation Overview Determining How to Mount the TransducerPortable Mounting Alternate Transducers and Mounting MethodsAlternate Transducers and Mounting Methods Trolling Motor MountingTransducer Exchange Beginning InstallationTransducer Exchange HOW Sonar Works Using the 250DXHOW Sonar Works Simulator Operation Feature MemorySimulator Operation What YOU see ON-SCREEN What You See On-ScreenDepth of the water changes, the range What YOU see on Screen Control Functions Control FunctionsUsing the 250DX Using the 250DX Menu Functions Menu FunctionsUsing the 250DX Depth Range Upper Zoom Limit To disable Zoom, press the Menu button until Fish symbols shown on-screen Using the 250DX Using the 250DX Using the 250DX Numeral Size. Two options are available Using the 250DX Using the Diagnostic Feature Advanced FeaturesUsing Diagnostic Maintenance MaintenanceMaintenance Nothing happens when I turn the unit on TroubleshootingTroubleshooting There is no transducer detected There is no bottom reading visible on the displayUnit comes on before I press POWER, and won’t turn off Get gaps in the reading at high speedsMy unit loses power at high speeds Screen begins to fade out. Images are not as sharp as normalCustomer Support Customer SupportSpecifications TOP