ON again and carefully turn the control knob until you feel a tingling or slight twitch under or around the electrodes.) Always start with the low- est setting and increase the intensity slowly.
If the stimulation levels are uncomfortable or become uncomfortable, reduce the stimulation intensity to a comfortable level; or cease stimula- tion and contact your physician.
6.Setting the Patient Compliance Counter:
a)To turn on the Patient Compliance Counter: While the unit is ON, hold down the UP key and press the MODE key simultaneously.
b)To reset the Patient Compliance Counter: While the unit is ON, press the UP key and the MODE key simultaneously (this will take you into the Compliance Counter), then push the DOWN key and press the MODE key simultaneously.
c)Press the UP and MODE keys simultaneously to return to the treat- ment status.
BATTERY INFORMATION
A
Replacing the Battery
When the LCD low battery mark illuminates, and the unit does not remain illu- minated once turned on, the battery should be replaced.
1.Turn unit OFF.
2.Remove the front panel cover by pressing on the top of the panel and pressing down in order to slide the panel down. Continue sliding the panel downwards until the panel is completely removed from the unit. This will reveal the battery compartment.
3.Remove the discharged battery from the device.
4.Place new battery in the compartment. Note: Be sure the proper polarity (+ and
11