| ERROR DISPLAYS AND TROUBLESHOOTING |
|
|
|
ERROR SYMOL / SYMPTOM | CAUSE / CHECK POINT | REMEDY |
|
|
|
No display when MEMORY BUTTON | Is there any measured result stored in the | Nothing is displayed when there is no |
( 3 or 4 )is pressed. | memory? | measured data stored. |
| Are you pressing the right MEMORY BUTTON? | Check the memory bank number. |
| Have the batteries run down? | Replace them with new ones. |
|
|
|
Cannot make great number of | Are you using alkaline batteries? | Use alkaline batteries to increase battery life. |
measurements. |
|
|
|
|
|
Cannot complete the measurement. | Have the batteries run down? | Replace all four batteries with new ones. |
|
|
|
Blood pressure is different each | Are you measuring with correct posture? | Measure with the correct posture. |
time. The reading is extremely low | Blood pressure readings constantly vary with | See TIPS ON TAKING YOUR BLOOD PRESSURE. |
(or high). | time of measurement and nervous condition. |
|
|
|
|
Pulse rate is too low (or too high). | Did you move during the measurement? | Remain still during the measurement. |
| Did you make measurement right after exercise? | Measure again after resting for more than 5 |
|
| minutes. |
|
|
|
The power is automatically turned | Have you left the instrument untouched after | This does not indicate a problem. The instrument |
off. | the measurement? | automatically turns off 3 minutes after a |
|
| measurement or 30 seconds after memory display. |
|
|
|
If you cannot get correct measurement with the methods above, do not tamper with the internal mechanism. Contact your dealer. If the unit malfunctions, please return it to the dealer or an authorized service representatives for service according to the warranty. Service data are provided to qualified dealers by the manufacturer.
- 16 - | - 17 - |