4.If you decide to abort the delete action instead, press to go back to the main screen.
Deleting All Records |
| |
1. Press | or | to recall the latest measurement |
record for User 1 or User 2.
2.Press and hold SET until the LCD shows “dEL ONE”.
3.Press to show the option of deleting your entire measurement history; the LCD displays “DEL ALL”.
4.Press SET to delete all measurement records. After you have confirmed deleting all records, the screen shows dEL dONe.
5.If you decide to abort the delete action instead, press
once OR press | twice to go back to the |
main screen. |
|
Troubleshooting
This section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If the product is not operating as you think it should, check here before arranging for servicing.
Problem | Symptom | Check This | Remedy | |
No | Display is | Batteries are | Replace with new | |
power | dim or will | exhausted | batteries | |
| not light up |
|
|
|
|
| Batteries | Insert the | |
|
|
| ||
|
|
| are inserted | batteries correctly |
|
|
| incorrectly |
|
|
|
|
| |
Low | shows | Batteries | Replace | |
batteries | on the | are low | with new | |
| display |
| batteries | |
|
|
|
|
|
Problem | Symptom | Check This | Remedy |
Error | Err shows | A measure- | Relax for a |
message | on the | ment error | moment and then |
| display. | occurred | measure again |
|
|
|
|
| Err 1, 2, or | The wrist cuff | Refasten the cuff |
| 3 shows | is not secure | and then measure |
| on the |
| again |
| display |
|
|
|
|
|
|
| Err 4 | The monitor | Movement |
| shows on | detected | can affect the |
| the display | motion while | measurement. |
|
| measuring | Relax for a |
|
|
| moment and then |
|
|
| measure again |
|
|
|
|
| Err 5 | Pressure is | Relax for a |
| shows on | over 37.3 kPa | moment and then |
| the display | (280 mmHg) | measure again |
|
|
|
|
| Err 6 | Deflation | Movement |
| shows on | period was too | can affect the |
| the display | long | measurement. |
|
|
| Relax for a |
|
|
| moment and then |
|
|
| measure again |
|
|
|
|
Problem | Symptom | Check This | Remedy |
Error | Err 7 or 8 | A calibration | Retake the |
message | shows on | error occurred | measurement. If the |
| the display |
| problem persists, |
|
|
| contact the retailer |
|
|
| or our customer |
|
|
| service department |
|
|
| for further |
|
|
| assistance. Refer |
|
|
| to the warranty for |
|
|
| contact information |
|
|
| and return |
|
|
| instructions |
|
|
|
|
Settings | Date and | The clock was | Reset the clock |
are wrong | time are | not set or reset |
|
| incorrect | after installing |
|
|
| new batteries |
|
|
|
|
|
| Measure- | The | Reset the |
| ment unit | measurement | measurement unit |
| (kPa or | unit was not set |
|
| mmHg) are | or reset after |
|
| incorrect | installing new |
|
|
| batteries |
|
|
|
|
|
13 | 14 |