General
The OSP100 is the ideal tool for periodic process monitoring and for measuring temperature quickly and easily, by contact with the target or from distance. An optional laser pinpointing method simplifies the identification of the measuring surface area. The OSP100 is the essential tool for the prevention of a variety of temperature related equipment or process malfunctions. It can prevent serious damage to bearings, motors, valves, electrical switches and distribution conductors, cooling or steam traps, thermal insulation deterioration, hot spots, engine performances, extrusion heads, depleted refrigeration system, etc.
The instrument is rugged, accurate and compact for fast and easy periodic process monitoring, to control quality and/or supervise production phases of cooked and refrigerated foods.
Specifications
!Spectral band:
!Display: 3 ½ digit custom backlighted LCD plus labels for max, min, emissivity, Hold, °C, °F, low battery.
!Resolution: 0.1°C up to 199.9°C. 1°C otherwise.
!Response time: 500 ms
!Target pinpointing: sight dot or circular laser sighting (L and XL models only)
!Emissivity: adjustable from 0.30 to 1.00
!Auto Power Off: Automatic power Off if no key is pressed for more than 40 sec.
!Alarm: low and high alarm with visual and acoustic beeper
!Signal processing: °C/°F, average, hold, max, min (where applicable)
!Accuracy IR channel:
range from 23°C to f.s. : ±1% of reading or ±1°C whichever is greater
range from
!RTD channel: Pt100 IEC751 α385 (probe excluded) Input range: from
Accuracy: ±0.3% of rdg or ±0.3°C whichever is greater
!Laser sight::
Wavelength: 650nm, Beam diameter: 3mm, Beam divergence: <0.5mrad;
Laser indicator: asterisk on display
!FDA Classification: Class II, Complies with 21CFR Chapter 1, Subchapt.er J
!Safety Classification: Class 2
!Power supply: alkaline or rechargeable battery
!Battery life: 60 h (back light & laser off)
!Ambient temperature:
!Storage temperature:
!Dimensions and weight: 180x140x45 mm 380g nett
When alkaline batteries are used, DO NOT connect power module to input jack.
CE certifications
This instrument conforms to the following standards:
EN
EN
Operational Flow Chart |
|
| |
NORMAL MODE KEYPAD |
|
| |
| Laser pointer enable/disable | Set + - | °F |
|
| °C | |
|
|
| |
Probe | Display external probe/infrared measurements |
|
|
Lamp | Lock | HAL | LAL |
Backlight ON/OFF | e |
| |
Hold | Hold measurements in lower display. Refresh hold. |
| |
OSP100 uses a |
| ||
| High Quality custom display |
Min/Max | Circular switch between Average, Minimum or Maximum measurements on lower display. |
|
|
2nd |
|
|
| Activate 2nd function keys. |
| ||||||||
|
| + |
|
|
|
|
|
|
| ||||
2nd |
|
|
|
|
|
|
| Laser beam mode: Continuous / Flashing (economy mode). | |||||
|
|
|
|
|
| ||||||||
|
|
| + |
|
|
|
|
| = |
|
| ||
2nd |
|
| Probe | °C/°F | Switch engineering temperature units. | ||||||||
|
|
|
|
|
| + |
|
|
|
|
|
|
|
2nd |
|
|
|
| Hold | = | HI | Set mode for HIGH alarm. | |||||
|
|
|
| + |
|
|
|
|
|
|
|
| |
2nd |
|
|
| Min/Max | = | LO | Set mode for LOW alarm. | ||||||
|
|
|
|
| + |
|
|
|
|
| = |
|
|
2nd |
|
|
| Lamp | Set ε | Set mode for Emissivity. |
SET MODE KEYPAD
Increase digit. | Probe |
| 2nd | |
Decrease digit. | °C/°F |
|
| |
Lamp | Hold | Min/Max | ||
| ||||
Store value and return in normal mode. | Set ε | HI | LO | |
|
| Set Alarm |
Applications
FOOD (HACCP) To control and supervise production phases, handling, transport, and conservation of cooked and refrigerated foods.
AUTOMOTIVE To check engine, bearings, muffler, brakes, etc.
INDUSTRIAL & ELECTRICAL Bearings, motors, valves, switches, insulation, etc.
HVAC/R Heating & cooling systems, thermal insulation, radiators, ducts, etc.
Basic functions
To take a temperature reading, point the unit at the target you wish to measure. Push and keep the trigger pushed. The current temperature reading is displayed. Last reading before previous
The OSP100L and 100XL models can be supplied with a laser sighting with either a “laser dot” to read the target area center or with a “laser circle” to read the target area. Press the
[*]key to enable/disable the laser. The message “LAS EN” or “LAS DIS” will be displayed. When enabled, the laser can be switched on by pressing the trigger button for more than 1 sec. The laser will automatically switch off when the trigger button is released. The emission indicator symbol (*) will appear on lower display.
Press in sequence the [2nd] and [*] keys to change the laser beam mode between Continuous and Flashing (to increase battery expected life) mode.
Press the [Lamp] key to switch
Press in sequence the [2nd] and [°C/°F] keys to change the engineering units between °C and °F.
Enhanced functions
External probe operation. Connect the Pt100 resistance thermometer to the appropriate connector. Switch the unit on and press the [Probe] key. The “EXT” message will be displayed on the lower display for Pt100 readings.
Hold measurement. Press the [HOLD] key to freeze the displayed temperature. The hold value will be displayed in lower display. On the main display the temperature is that actually measured. Press the key again to refresh the Hold indication.
The Hold indication is refreshed by pressing the trigger for more than 1 sec with the unit in measuring mode and releasing the trigger. Switch the OSP100 off by pointing the unit to the target to recall the last value when you switch the unit on.
Select Average (AVG), Minimum (MIN) and Maximum (MAX) values. Press the [Min/Max] key until the appropriate message is displayed. The unit will refresh the values of each measurement from the power on. To reset the values, press the [2nd] key or switch the unit off and on.
Emissivity adjustments. Press in sequence the [2nd] and [Set ε] keys to change the emissivity from 0.30 to 1.00 with 0.01 resolution. On the lower display the set value is shown. Press the [UP] and [DOWN] keys to increase and decrease the emissivity. Press the [ENTER] key to store and return on the normal mode measurement.
Set low (LAL) and high (HAL) alarms. Press in sequence the [2nd] and [HI] keys to change the High alarm set point.. “HAL” will be seen on the display. The stored High alarm will be displayed. The High alarm level is to be set by adjusting the [UP] or [DOWN] keys. Press the [HI] key to enable/disable the acoustic alarm. Press the [ENTER] key to store the displayed value and return to normal mode. Press in sequence the [2nd] and [LO] keys to change the Low alarm set point..“LAL” will be seen on the display. The Low alarm level is to be set by adjusting the [UP] or [DOWN] keys. Press the [LO] key to enable/disable the acoustic alarm. Press the [ENTER] key to store the displayed value and return in normal operation mode. When the measurement is inside the programmed alarm band, the “LAL” or “HAL” symbol will be lit and an intermittent beep will be emitted.
Optics
Distance: Sensor to Object (in)
| 0 | 9 | 20 25 | 40 | 80 | 120 |
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 9.4 |
| |
DIA* (in) |
|
|
|
|
|
|
|
|
| 6.2 |
|
|
| |
|
|
|
|
|
|
| 3 |
|
|
|
|
| ||
Spot |
|
| 1 |
| 1.4 |
|
|
| 4.0 |
| ||||
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
| .5 |
|
|
|
|
| 2.7 |
|
|
| ||
|
|
|
|
|
| 1.4 |
|
|
|
|
| |||
|
|
|
|
| .9 |
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
| 23 | 35 |
|
|
|
|
| |||
|
|
| 12 |
|
|
|
|
|
|
|
| |||
(mm) |
|
|
|
|
|
|
| 67 |
|
|
| |||
25 |
| 35 |
|
|
| 100 | Nominal D:S = 30:1 | |||||||
|
|
|
|
|
|
|
|
|
|
|
| OSP100XL | ||
SpotDIA* |
|
|
|
|
|
|
| 75 |
|
|
|
|
| |
|
| * Spot size measured | 157 |
|
|
| ||||||||
|
|
| at 90% energy |
|
|
|
| 238 | OSP100, OSP100L | |||||
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
| Nominal D:S = 12:1 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| 0 | 225 500 625 | 1000 | 2000 | 3000 |
| |||||||
|
|
|
|
|
|
|
| Distance: Sensor to Object (mm) |
| |||||
Warnings and Cautions | OSP100L and OSP100LX models only |
!USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER THAN THOSE SPECIFIED HERE MAY RESULT IN HAZARDOUS RADIATION EXPOSURE.
!DO NOT LOOK AT THE LASER BEAM COMING OUT OF THE LENS OR VIEW DIRECTLY WITH OPTICAL INSTRUMENTS - EYE DAMAGE CAN RESULT.
!USE EXTREME CAUTION WHEN OPERATING THE LASER.
!NEVER POINT THE LASER BEAM AT A PERSON.
!KEEP OUT OF REACH OF ALL CHILDREN.
C A U T I O N | THIS APERTURE | ISEMITTED FROM | LASERRADIATION | AVOIDEXPOSURE | LASER RADIATION DO NOT STARE |
| INTO BEAM OR VIEW DIRECTLY | ||||
LASER RADIATION - DO NOT STARE INTO BEAM | WITH OPTICAL INSTRUMENTS. | ||||
MAX.OUTPUT<1mW, WAVELENGTH | MAX.OUTPUT<1mW, WAVELENGTH 630- | ||||
CLASS I I LASER PRODUCT | 670nm, CLASS 2 LASER PRODUCT | ||||
COMPLIES WITH 21CFR CHAPTER 1, SUBCHAPTER J | CONFORMS TO |
WARNING
DO NOT ATTEMPT TO OPEN THE LASER SIGHT MODULE.
(THERE ARE NO USER SERVICEABLE PARTS IN THE MODULE).
PATENT NOTICE: U.S. PAT. B1 5,368,392; 5,524,984; 5,727,880; 5,465,838; 5,823,678; 5,823,679. Other U.S. and Foreign Patents and Applications Pending. Manufactured in Italy. Licensed by Omega Engineering, Inc.