Invacare 7400036, 7400037 Converting from Patient Operated Wheel Locks Figure

Page 3

CONVERTING FROM PATIENT OPERATED WHEEL LOCKS (FIGURE 4)

NOTE: Refer to INSTALLATION WARNING in the SAFETY SUMMARY in this instruction sheet.

1.Remove the existing axle mounting screw and locknut that secures the rear portion of the patient operated wheel lock assembly, pivot arm spacer and rear wheel to the wheelchair frame.

2.Remove the 5/16-inch mounting screw, bushing, washer, and locknut that secures the strap and stud assmebly of the patient operated wheel lock to the wheelchair frame.

3.Secure the rear wheel to the wheelchair frame with the axle mounting screw and the 3/4-inch spacer provided. Securely tighten.

4.Install the attendant operated wheel locks. Refer to REPLACING THE ATTENDANT OPERATED WHEEL LOCKS in this instruction sheet.

STEP1

Locknut

Rear Portion of the Patient

Operated Wheel Lock

 

Axle Mounting

 

 

Screw

 

Wheelchair

 

 

Frame

 

 

Pivot Arm

 

 

Spacer

 

 

 

Rear Wheel

 

STEP2

STEP3

 

5/16-inch

 

 

Mounting Tube Clamp

Wheelchair

Locknut

Screw

Frame

 

Axle

Mounting

Screw

Thin

Washer

 

 

Spacer

 

 

Rear

 

Locknut

Wheel

Bushing

 

 

 

Strap and Stud Assembly of the

Wheelchair Patient Operated Wheel Lock

Frame

FIGURE4-CONVERTINGFROMPATIENTOPERATEDWHEELLOCKS

3

Image 3