10.FREQUENTLY ASKED QUESTIONS

For integrated CPAP devices:

a. When I wake up in the morning, my nose and throat feel dry. What can I do?

Try increasing your humidity setting. If this does not help, please contact your clinician for advice.

b. How do I prevent condensation in the tubing?

The humidity setting enables adjustment of humidity, so that there should be fine misting in the six inches (183cm) of the breathing tube closest to your face. If the humidity setting is too high for the conditions, condensation may occur in the breathing tube.

Note: Using greater than 6' (183 cm) of tubing will increase condensation.

There are several ways of reducing condensation in the tube:

iEnsure the device is not positioned in a cool draft.

iiDecrease the humidity setting on your device slightly. Gradually decrease the setting until the condensation no longer occurs.

iiiWarm the air in the room.

c. Do I have to use distilled water with my device?

The use of distilled water will maximize the life of the water chamber. Water from the faucet will often contain minerals which can damage the chamber, causing pitting in the base, even if the water is passed through a filter.

d. When do I replace my water chamber?

Replace the chamber if the plastic walls of the chamber become cracked or cloudy or the chamber base becomes pitted. (Note: If the chamber leaks at all, it should be replaced immediately).

Can I use my device in other countries?

Yes. Simply use the appropriate electrical socket adaptor and the device will automatically adapt to any voltage supply.

Note: When moving your device, ensure that the water chamber is empty. Machine failure due to water damage is not covered by warranty.

Can I use supplemental oxygen with my device?

Yes, oxygen can be administered at the mask. Turn the device on before turning on the oxygen. Ensure that the oxygen source is turned off before turning off the device to avoid oxygen accumulating in the machine.

English

11.SPECIFICATIONS

DIMENSIONS:

275mm x 170mm x 140mm

 

(10.8" x 6.7" x 5.5")

WEIGHT:

1.9kg (4.2lbs)

 

(3.0kg packaged in bag incl. accessories)

ELECTRICAL RATINGS

 

Supply Frequency:

50-60 Hz

Supply Voltage and Current:

100-115V~ (1.2A, 1.3A max),

 

220-240V~ (0.8A, 1.8A max)

 

Note: These values represent

 

average current

DC to AC Inverter requirements: 115V - 200W/300W surge

 

230V - 300W/500W surge

Heater Plate :

85W max

Heater Plate Temperature:

65°C (149°F) max

PERFORMANCE

 

Pressure Range

4 to 20 cmH2O

Altitude Range

0 to 3000m, or 0 to 9000 ft

STANDARDS COMPLIANCE Designed to conform to the

requirements of:EN / IEC 60601-1 UL60601-1

CSA C22.2/No.601.1 AS3200.1.0

The device complies with the electromagnetic compatibility requirements of IEC 60601-1-2. In certain circumstances, the device may affect or be affected by nearby equipment due to the effects of electromagnetic interference. If this should happen, try moving your unit or the location of the device causing interference, or alternatively consult your homecare provider.

A - 9

Image 13
Contents Operating Manual Page Package Contents Page SleepStyle 200 Series Oxygen USE Warnings Symbol DefinitionsIntended USE Important Parts of Your Device Integrated Range Description of the DeviceSetup Instructions Cleaning & MaintenanceDisplay Description and Functions Controls and Display Integrated Cpap RangeDisplay Patient Menu Summary Integrated Cpap Range23x Controls and Display Standard Cpap RangePatient Menu Summary Standard Cpap Range Patient Menu Additional Settings Summary Standard Cpap RangeSpecifications Frequently Asked QuestionsOperating Conditions TroubleshootingStorage and Transport Conditions AppendixREV F