Troubleshooting

Problem/possible Solution cause

Mask is uncomfortable

Cushion incorrectly

With the air turned off, adjust the straps so that you can

inflated

insert two fingers between your cheek and the lower

 

headgear strap.

Technical specifications

Dead space

The dead space of the mask varies according to cushion

information

size. It is 185 mL for the Medium size.

Resistance

Drop in pressure measured (nominal)

 

at 50 L/min: 0.1 cm H2O

 

at 100 L/min: 0.7 cm H2O

Gross dimensions

Mask fully assembled – no headgear.

S,

5.55”/141 mm (H) x 3.58”/91 mm (W) x 4.21”/107 mm (D)

M, L, LW:

6.06”/154 mm (H) x 3.74”/95 mm (W) x 4.6”/117 mm (D)

Mask setting

Select ‘Activa’ (if available) or ‘Standard’ as the mask option

options

when using the Mirage Activa LT with ResMed devices that

 

have mask setting options.

2

Image 16
Contents Mirage SoftGel Mirage TM SoftGel Mirage SoftGel Iii 61600 S 61602 61601 M 61603 Disassembly / Démontage / Desmontaje / Desmontagem Reassembly / Remontage / Montaje / Remontagem Uso previsto Utilização da máscaraLimpeza da máscara em ambiente domiciliar Diariamente/Após cada utilizaçãoPrecaução Reprocessamento da máscara entre pacientesResolução de problemas Especificações técnicas PortuguêsArmazenamento DescarteSímbolos Garantia limitadaPage Mirage Activa LT Intended useTroubleshooting Technical specificationsMédio 789 Manufacturer Distributed by Mirage SoftGel Amer pOr