5.6Temperature

5.6.1The temperature of the process can be read at any time by entering the menu and calling for TEMPERATURE. Either °C or °F will be indicated depending on the position of DIP switch No. 1. Bank S1. See Section 4.8.

5.6.2The 0-5 Vdc and 0-1 mA analog outputs can be dedicated to follow the process temperature by simply placing DIP switch No. 8 in the off position. The temperature span of the output is set to the utility menu. See Section 4.8 and 7.4.

5.7Status

5.7.1The DOCN600 continuously checks the integrity of all stored data and monitors the condition of the measuring system. If a fault is detected, the FAIL LED above the display will turn red. The STATUS in the operation menu will provide a numerical code, giving a possible cause and a suggested remedy.

5.7.2The following table shows the display codes, causes and remedies:

Code

Possible Cause

Suggested Remedy

0

Normal Condition

No action required

1

Reserved

 

2

Temperature Sensor off Scale

Verify process. Check for open or short

 

 

connections.

3

D.O. reading off Scale

Verify process. Check for open or short

 

 

connections.

4*

Memory Loss

Call your OMEGA Rep. Or OMEGA

 

 

directly

5

Reserved

 

6

Sensor Fault

Check probe membrane for perforation

7

Factory Setting in force, as a

Perform procedure according to Section 9.2

 

result of the ESCAPE procedure

 

NOTE: Code 4 could be a serious failure so the alarm relay will energize in addition to the red illumination of the FAIL LED, if DIP switch 5 Bank S1 is OFF.

5.8Relay A Setpoint

5.8.1As shipped from the factory, Relay A is configured to control decreasing D.O. However, you may change the direction of control by changing the position of DIP switch 4 Bank S1. Please refer to Section 4.8.

Page 15

Image 18
Contents DOCN600 Servicing North America Table of Contents General Information Specifications Power RequirementsModel DOCN600 Additional Features Type of Mounting InstallationLocation Conduit ConnectionsSensor Installation, Flotation Mounting Page Sensor Installation, Submersion Mounting Probe ConnectionsExtension Cable Junction Box Description of Functions OverviewCalibration Temperature CompensationMembrane Alarm Relay Relay OutputsAnalog Outputs Operation MenuDIP Switches OFFOutput Hold Simulated Input for TestingWatchdog Timer and Self Diagnostics Parameter and Operation CheckingPassword START-UP and OperationPercent Saturation/ppm Analog Output Range Expand TestTemperature StatusRelay a Setpoint Relay a Deadband Relay B SetpointRelay B Deadband Operating Hints Alarm HighAlarm Low Cell CareUtility Menu ResetAccess to Utility Menu Utility Menu FunctionsTemperature Output Adjust 0-1 mA / 0-5 Vdc OutputAdjust 4-20 mA Output Description DiagnosticsAltitude Compensation Invalid OutputTroubleshooting and Service Isolate the causeEscape Printed Circuit Board Replacement Sensor FaultRelay Replacement Statements of Conformity from the Manufacturer WARRANTY/DISCLAIMER Shop online at omega.com

DOCN600, Dissolved Oxygen System specifications

Omega Engineering has set a standard in the field of environmental testing and monitoring with its advanced Dissolved Oxygen System, the DOCN600. This sophisticated device is engineered for precision and reliability, making it a preferred choice for various applications, from wastewater treatment to aquaculture, and laboratory settings. The DOCN600's primary mission is to accurately measure dissolved oxygen levels, a critical parameter in assessing water quality and supporting aquatic life.

One of the standout features of the DOCN600 is its state-of-the-art optical dissolved oxygen sensor technology. Unlike traditional Clark-style electrochemical sensors, which can be subject to interference and require frequent calibration, the optical method uses luminescence to measure oxygen levels. This technology offers several advantages, including lower maintenance requirements, enhanced stability, and improved accuracy over a broad range of conditions.

The DOCN600 boasts a user-friendly interface, complete with a high-resolution LCD display that provides real-time data visualization. Users can access multiple display modes, allowing for quick reference of key metrics, including temperature and oxygen saturation levels. Furthermore, the system is designed with intuitive menus that facilitate easy navigation, ensuring that both seasoned professionals and beginners can operate the device with confidence.

For data management, the DOCN600 is equipped with advanced connectivity options, including USB and RS-232 ports. This allows for seamless integration with external devices and enables users to export data for further analysis. Coupled with its internal memory, which can store extensive historical data, this system empowers users to monitor trends over time, aiding in decision-making processes.

Another notable characteristic of the DOCN600 is its robust build quality, ideal for deployment in various environmental conditions. The system is engineered to be resistant to water, dust, and harsh chemicals, ensuring longevity and reliable performance even in challenging environments. Additionally, with its compact and portable design, it can be conveniently used in field applications as well as in stationary setups.

In summary, the Omega Engineering DOCN600 Dissolved Oxygen System combines cutting-edge optical technology, user-friendly features, advanced connectivity, and a durable design. It stands out as an invaluable tool for anyone needing precise and reliable dissolved oxygen measurements. Whether used in research, environmental monitoring, or industrial applications, the DOCN600 is an excellent investment for achieving accurate water quality assessments.