medicalManual.RevD.qxd 9/9/04 4:18 PM Page 8
1.2 Component Identification
FIGURE 1
1.3Component Description
1.3
2.Low Battery Indicator - The low battery indicator is located at the top of the display and is only activated when the voltage on the batteries is below a normal operating level.
3.“%” symbol - The “%” sign is located to the right of the concentration number and is present during normal operation.
4.Calibration symbol - The calibration symbol is located at the bottom of the display and is timed to activate when a calibration is necessary.
5.ON/OFF Key - This key is used to turn the device on or off.
6.Calibration Key - This key is used to calibrate the device. Holding the key for more than three seconds will force the device to enter a calibration mode.
CAUTION: The device will assume a percent oxygen concentration when calibrating. Be sure to apply 100% oxygen, or ambient air concentration to the device during calibration or the device will not calibrate correctly.
Sample Inlet Connection
This is the port at which the device is connected to determine oxygen concentration.
1.4 MAX-250+ Oxygen Sensor
The MAX-250+ oxygen sensor offers stability and extra life.
The
CAUTION: The Maxtec®
CAUTION: Do not use ethylene oxide sterilization. Do not immerse the sensor in any cleaning solution, autoclave or expose the sensor to high temperatures.
CAUTION: Dropping sensor can adversely affect its performance.
2.0OPERATING INSTRUCTIONS
2.1Getting Started
2.1.1 Protect Tape
Prior to turning on the unit, a protective film covering the threaded sensor face must be removed. After removing the film, wait approximately 20 minutes for the sensor to reach equilibrium.
2.1.2 Remove the Battery Ribbon
A ribbon has been inserted between the two case halves to prevent a battery connection. Remove the ribbon by pulling it completely out of the case. To energize the unit tighten all three screws with the included Phillips screwdriver.
NOTE: If you do not tighten all three screws, the unit may not turn on or it may erratically turn on and off.
2
M A N U FA C T U R E D B Y M A X T E C , I N C . | W W W. M A X T E C I N C . C O M |
3