Graham Field JB0150-085, JB0150-086 warranty Oxygen Flow Control Regulators, Set-Up Yoke Style

Page 2

Oxygen Flow Control Regulators

Oxygen flow control regulators are pressure-reducing devices which lower oxygen pressure from a cylinder to levels that can be safely used by the end user.

Oxygen flow control regulators are intended for the admin- istration of oxygen to patients who need increased oxygen levels to improve their breathing conditions. In addition, some models may be used in emergency situations. In an emergency situation, be sure to contact a physician or EMS immediately.

WARNING: DO NOT reuse plastic crush gaskets (fire risk).

Set-Up (Yoke Style)

Inspect the cylinder following the instructions of your gas supplier. Be sure to purge the cylinder valve seat of debris.

Inspect the regulator for oil or grease. If either is vis- ible, DO NOT use the regulator.

Be sure the sealing washer is in place on the inlet of the regulator.

Before attaching the regulator to the valve, inspect the regulator and the seal to ensure that they’re in good condition, and be sure that the regulator has just one of the CGA-870 seals. Slip the yoke over the cylinder post and fit the pins into the holes on the cylinder valve. The pins should fit easily into these holes.

Tighten the T-handle firmly by hand, but don’t use wrenches or other tools that may over-torque the handle. Turn the T-Handle clockwise until the screw is seated in the dimple on the cylinder valve. Be sure the regulator is tightened securely into the cylinder.

Attach the oxygen supply tubing to the outlet connec- tion of the regulator.

Image 2
Contents Oxygen Regulator Oxygen Flow Control Regulators Set-Up Yoke StyleSet-Up H-Nut Style Instructions for UseLimited Lifetime Warranty