set up & operation | 7 |
|
BATTERY OPERATION: To ensure proper operation from internal battery, fully charge the battery for 1.5 hours as explained in the Battery Charging section. To operate the unit from an internal rechargeable battery, ensure that no external power sources are plugged into the DC power input on the side of the unit.
Once the power source is selected, simply turn the unit on using the power switch located on the side of the unit. The power indicator light (displayed on the front of the unit) will glow yellow to show which power source is being used and glow green when an external power is connected.
battery charging/ cleaning
IMPORTANT OPERATION AND BATTERY INSTRUCTIONS:
Your 18605 portable suction machine is capable to run on AC or DC power. To ensure full func- tion of this unit, the battery must be completely charged before the first use. Do not turn on the unit before the battery is completely charged. It takes approximately 1.5 hours to fully charge the unit. Charge the unit when battery levels are low to ensure proper operation.
WARNING
If the unit does not receive an external power source or the battery does not get recharged im- mediately, the low battery indicator light will remain on and the performance of the unit will drop off rapidly. Switch to another power source immediately to avoid an interrupted suction procedure.
Adjust the vacuum level from 0 to 600 mm Hg by turning the vacuum regulator knob located on the side of the unit (clockwise to increase vacuum and
NOTE
Gauge is for reference only. If the unit sustains a severe drop, accuracy of the gauge must be checked.
Suction at the patient is automatically obstructed when liquid level reaches the float
NOTE
Always transport unit with vacuum regulator knob rotated fully clockwise in case unit is dropped.
CAUTION
When automatic float
CAUTION
Should fluid be aspirated back into the unit, equipment provider servicing is necessary as pos- sible vacuum pump damage may result.