Error Codes
Error codes are shown in the main display for approx. 2 seconds, after which the program automatically returns to the weighing mode.
Display | Cause | Solution |
No segments appear on | No AC power is available |
the display | The AC adapter is not |
| plugged in |
| Battery is dead |
Check the AC power supply Plug in the AC adapter
Replace the battery Recharge the battery using an external recharging device
h | The load exceeds the | Unload the scale |
| scale capacity |
|
|
|
|
l | The load plate is not in place | Place the load plate on the |
| Something is touching | scale |
| Move the object that is | |
| the load plate | touching the load place |
|
|
|
e 01 | Display capacity exceeded: | Decrease the weight on |
| Value to be output cannot be | the scale |
| shown on the display |
|
|
|
|
e 02 | Calibration parameter | Calibrate only when zero |
| not met; e.g.: | is displayed |
| – scale not zeroed | Press t to tare the |
|
| scale |
| – scale is loaded | Unload the scale |
|
|
|
e 09 | When gross - zero, no tare | Tare the scale |
|
|
|
e 10 | The t key is blocked |
| when there is data in |
| the second tare memory |
Press [CAL/CF] > 2 sec. to clear the tare memory and release the tare key
e 11 | Value input is not allowed for | Press t |
| second tare memory |
|
|
|
|
e 22 | Weight is too light or there is | Increase the reference quantity |
| no sample on the balance |
|
|
|
|
e 30 | Interface port for printer | Contact your local |
| output is blocked | Service Center |
|
|
|
Max. weighing range is less than indicated under ”Specifications”
The scale was
turned on without the weighing pan in place
Place the weighing pan on on the scale
and press [ON/OFF] to turn on the scale
The weight readout | The scale has not | Calibrate/adjust the scale |
is obviously wrong | been calibrated/adjusted | Tare before weighing |
| The scale was not | |
| tared before weighing |
|
|
|
|
If any other errors occur, contact your local Service Center!