A&D HJ-150 specifications Automatic power-off function, Scale

Page 3

Automatic power-off function

The automatic power-off function turns power off after approximately 2 minutes of stable display during the object(s) applied. When there is no object applied, it will turn off after 30 seconds.

Calibration

1. Make sure that there is nothing on the weighing pan, press and hold the LIGHT/CAL key approximately 3 seconds. The display shows

Cal . Release the key.

2.The scale memorizes the zero point and automatically displays

100.0 g .

3.Place the calibration weight of 100g on the center of the pan within 10 seconds.

4. After approximately 8 seconds, the scale will display f for success and e for failure. Then, the scale automatically turns off.

5. Turn the scale on again to start the operation.

G If calibration fails, check if anything is touching the weighing pan, there is vibration or wind, and the calibration weight is correct.

Dimension

(mm)

20.3

60

102

64

98

 

 

 

70

 

66

 

 

 

 

 

 

 

 

 

 

 

Scale

Protective cover

 

3

Image 3
Contents Specifications BatteriesCapacity 150 x 0.1 g / 5.29 x 0.01 oz Display Weighing procedures Installing / Replacing batteriesScale Automatic power-off functionKorea Limited German Sales Office