Omron Healthcare 510 Specifications, Measured in the left ear?, Temperature shown is rather low

Page 13

GentleTemp 510 English

Is the temperatures measured in the right ear different from that

Battery life:

With a new battery approx. 5000 measurements

measured in the left ear?

 

Sensor:

Thermopile

Among healthy people, there should be no significant difference in the

Temperature indication:

In 3 digits of +°C, every 0.1°F or °C

measurement results. Differences may be caused by the following reasons:

Precision:

According to NEN-EN 12470- 5 Clinical

1) The infrared sensor is not inserted in the same way.

(Celsius)

thermometers – Part 5 Performance of infra-red

2) Measurement is not conducted by inserting the unit in a stable manner in

 

ear thermometers (with maximum device)

the same angle. Try to measure in either ear that can show the highest

 

34°C – 42.2°C: ±0.2°C

measurement value constantly. In accordance with the “10 seconds mode,”

 

Based on the measurement of the standard black

the unit detects the highest temperature. (Refer to paragraph 3.2).

Precision:

body at the room temperature of 25°C (RH 50 %)

 

 

 

According to NEN-EN 12470- 5 Clinical

The temperature shown is rather low.

(Fahrenheit)

thermometers – Part 5 Performance of infra-red

1

The probe cover is dirty.

 

 

ear thermometers (with maximum device)

2

The infrared sensor is dirty.

 

93.2°F – 108°F: ±0.3°C

3

You removed the unit from the ear before the measurement is finished.

 

Based on the measurement of the standard black

4

The ear is cold. The temperature tends to indicate low when you use

Measurement range:

body at the room temperature of 77°F (RH 50 %)

 

an ice bag or an ice pack, or immediately after coming in from the

34°C - 42.2°C (93.2 - 108°F)

5

cold in winter.

 

Weight:

Approx. 50 g (1.8 oz) (including the battery)

The thermometer is not inserted deep enough in the ear.

External dimensions:

Approx. 4.6 cm (B) X 9.3 cm (H) X 5.7 cm (D)

6

The infrared sensor is not directed towards the eardrum.

Operating environment:

Ambient temperature; 10°C to 40°C (50°F to 104°F)

 

Use the 10 seconds measurement mode (Refer to paragraph 3.2).

Relative humidity:

30 - 85%

6

SPECIFICATIONS

Storage environment:

Ambient temperature; -20°C-60°C (-4°F- 140°F)

Relative humidity:

10 - 95%

Name:

GentleTemp 510®

Accessories:

Storage case, Instruction Manual, probe covers

 

(10 pieces)

Model:

MC-510-E2

Options:

A probe cover is already attached to the main unit.

Voltage:

3V DC (lithium battery CR2032)

Probe cover exclusively for GentleTemp 510®

Power consumption:

0.05 W

 

series, Probe Cover MC-EP1

13

Image 13
Contents GentleTemp GentleTemp 510 English Table of ContentsGeneral Safety precautionsStore the unit out of children’s reach Correct measurement results may not be obtainedCare and storage General AdviceComponents of the thermometer Measuring the temperature of an infant Clean probe coverPreparation Before Measurement Move the unit in the earRecommendations To change the temperature from C to FUsage of the Thermometer Mode 1 the fast measurement 1- 3 seconds in 7 stepsMode 2 the 10 seconds measurement in 9 steps To finish the measurement To measure repeatedly Battery replacement Normal and raised temperature Temperature measurements taken within the ear To Interpret Measured Results Age Normal ear temperature in C and F Solving ProblemsEar temperature compared to other types of body temperature Troubleshooting Display What to check? How to correct?Temperature indicated is rather high Temperature shown is rather low SpecificationsMeasured in the left ear? Production facility EU-representativeSubsidiary Handelsgesellschaft mbH