GentleTemp 510 English
Is the temperatures measured in the right ear different from that | Battery life: | With a new battery approx. 5000 measurements | ||
measured in the left ear? |
| Sensor: | Thermopile | |
Among healthy people, there should be no significant difference in the | Temperature indication: | In 3 digits of +°C, every 0.1°F or °C | ||
measurement results. Differences may be caused by the following reasons: | Precision: | According to | ||
1) The infrared sensor is not inserted in the same way. | (Celsius) | thermometers – Part 5 Performance of | ||
2) Measurement is not conducted by inserting the unit in a stable manner in |
| ear thermometers (with maximum device) | ||
the same angle. Try to measure in either ear that can show the highest |
| 34°C – 42.2°C: ±0.2°C | ||
measurement value constantly. In accordance with the “10 seconds mode,” |
| Based on the measurement of the standard black | ||
the unit detects the highest temperature. (Refer to paragraph 3.2). | Precision: | body at the room temperature of 25°C (RH 50 %) | ||
|
|
| According to | |
The temperature shown is rather low. | (Fahrenheit) | thermometers – Part 5 Performance of | ||
1 | The probe cover is dirty. |
|
| ear thermometers (with maximum device) |
2 | The infrared sensor is dirty. |
| 93.2°F – 108°F: ±0.3°C | |
3 | You removed the unit from the ear before the measurement is finished. |
| Based on the measurement of the standard black | |
4 | The ear is cold. The temperature tends to indicate low when you use | Measurement range: | body at the room temperature of 77°F (RH 50 %) | |
| an ice bag or an ice pack, or immediately after coming in from the | 34°C - 42.2°C (93.2 - 108°F) | ||
5 | cold in winter. |
| Weight: | Approx. 50 g (1.8 oz) (including the battery) |
The thermometer is not inserted deep enough in the ear. | External dimensions: | Approx. 4.6 cm (B) X 9.3 cm (H) X 5.7 cm (D) | ||
6 | The infrared sensor is not directed towards the eardrum. | Operating environment: | Ambient temperature; 10°C to 40°C (50°F to 104°F) | |
| Use the 10 seconds measurement mode (Refer to paragraph 3.2). | Relative humidity: | 30 - 85% | |
6 | SPECIFICATIONS | Storage environment: | Ambient temperature; | |
Relative humidity: | 10 - 95% | |||
Name: | GentleTemp 510® | Accessories: | Storage case, Instruction Manual, probe covers | |
| (10 pieces) | |||
Model: | Options: | A probe cover is already attached to the main unit. | ||
Voltage: | 3V DC (lithium battery CR2032) | Probe cover exclusively for GentleTemp 510® | ||
Power consumption: | 0.05 W |
| series, Probe Cover |
13