Fluke PN 3612805 user manual Cat

Page 10

373

Users Manual

Symbol

Meaning

Symbol

Meaning

 

 

 

 

 

IEC Measurement Category III

 

IEC Measurement Category IV

 

CAT III equipment has protection

 

CAT IV equipment has protection

CAT III

against transients in equipment in

CAT IV

against transients from the primary

fixed-equipment installations, such

supply level, such as an electricity

 

as distribution panels, feeders and

 

Meter or an overhead or

 

short branch circuits, and lighting

 

underground utility service.

 

systems in large buildings.

 

 

,

Application around and removal

 

 

from HAZARDOUS LIVE

 

 

 

conductors is permitted.

 

 

 

 

 

 

Note

The Measurement Category (CAT) and voltage rating of any combination of test probe, test probe accessory, current clamp accessory, and the Meter is the LOWEST rating of any individual component.

6

Image 10
Contents 373 Limited Warranty and Limitation of Liability Table of Contents 373 Introduction Safety Information Clamp Meter 373 Symbol Meaning CAT Meter Auto Off Backlight Ghl003.eps Or V 30 Ghl005.eps Hold Ghl007.eps Electrical Specifications Mechanical Specifications AA, Neda 15A, IEC LR6 373