Fluke 80T-150UA instruction sheet Specifications, Electrical, Environmental

Page 2

Specifications

The 80T-150UA will achieve rated accuracy when it is used with any 0.25 % DMM that has an input impedance of 1 MΩ.

Electrical

Measurement Range: (See Probe Limitations) -50 to +150 °C

-58 to +302 °F

Accuracy: (Applies for one year after purchase or calibration)

AMBIENT °C

ACCURACY

+15 to +35

°C

±1 °C from 0 to +100 °C, decreasing

 

 

linearly to ±3 °C at -50 and +150 °C

0 to +15 °C and

±2 °C from 0 to +100 °C,

+35 to +50

°C

decreasing linearly to ±4 °C at -50

 

 

and +150 °C

AMBIENT °F

ACCURACY

+59 to +95

°F

±1.8 °F from +32 to +212 °F,

 

 

decreasing linearly to ±5.4 °F at -58

 

 

and +302 °F

+32 to +59

°F and

±3.6 °F from +32 to +212 °F,

+95 to +122 °F

decreasing linearly to ±7.2 °F at -58

 

 

and +302 °F

Sensitivity (80T-150UA output): 1 mV dc / °C or °F

Voltage Standoff: 60 V dc or 30 V ac rms (42.4 V peak)

Settling Time: 9.5 seconds to settle within 2 ° for a 50 ° change

RF Field: The 80T-150UA product is unspecified in environments with EMC fields greater than 100 mV/meter. ESD performance is specified as Criteria C per EN61326.

Environmental

Ambient Operating Range for Temperature Module: 0 to +50 °C (+32 to +122 °F)

Maximum Temperature Probe Body and Cable: +70 °C (+160 °F) See Probe Limitations

Minimum Temperature Probe Body and Cable:

Minimum Operating Temperature for Cable: -12 °C (+10 °F)

Minimum Operating Temperature for Probe Body: -40 °C (-40 °F) See Probe Limitations

Storage Temperature for Unit: -40 to +70 °C (-40 to +160 °F)

Humidity: 0 % to 90 % (0 °C to +35 °C)

0 % to 70 % (+35 °C to +50 °C) (noncondensing)

Altitude: Operating: 3048 m ( 10,000 ft)

Storage: 15240 m (50,000 ft)

Application Force: 15 pounds maximum (probe tip to measured surface)

Image 2
Contents Symbols IntroductionEnvironmental SpecificationsElectrical Sensitivity 80T-150UA output 1 mV dc / C or FGeneral Error SourcesOperating Notes Probe LimitationsBattery Condition Test General MaintenanceAccess Information OperationPerformance Test CleaningBattery Replacement CalibrationLimited Warranty & Limitation of Liability Test Equipment Requirements