Fluke 80T-IR, E instruction sheet Safety Information

Page 2

Safety Information

The 80T-IR/E complies with IEC Publication 1010-1- 1990 including Amendment 1, CSA C22.2 No. 231, ANSI/ISA-S82.01 and .03 Safety Standards.

WWARNING

IF TARGET EMMISSIVITY IS LESS THAN 0.95, THE PROBE CAN INDICATE A TEMPERATURE LOWER THAN THE ACTUAL TARGET TEMPERATURE. AVOID TOUCHING THE TARGET; THERMAL BURNS COULD RESULT.

CAUTION

Do not place the probe on or around hot objects (70°C / 158°F). It will damage the probe case.

If the probe is exposed to significant changes in ambient temperature (hot to cold or cold to hot), allow 20 minutes for temperature stabilization, before taking measurements.

Do not operate the probe near large electrical or magnetic fields such as arc welders and induction heaters. These fields can cause measurement errors.

Condensation may form on the lens when going from a cold to hot environment - wait 10 minutes for condensation to dissipate before taking measurements.

Connectors must only be plugged into voltage measurement input jacks of the test instruments.

Do not touch or hold by the front cone. Temperature readings can be affected by heat from the hand.

Equipment use not specified by manufacturer may impair safety.

Image 2
Contents Box Contents 80T-IR/ESafety Information Compatibility OperationDisplay Codes EmissivityDistance to Spot Size Ratio Display Codes Condition ActionDiameter of spot Quick Check SpecificationsMeasurement Considerations Case Cleaning Battery ReplacementFront-Window Cleaning as necessary Limited Warranty & Limitation of Liability Calibration ServiceReplacement Part