Fluke 07-402 manual Introduction, Section

Page 5
Introduction 1

Introduction 1

Introduction

Section 1

Introduction

1.1 Introduction

The Portable Digital Thermometer provides a rapid and reliable means of checking solution temperatures in x-ray film processors. It features a detachable temperature-sensing probe with a 3-foot coiled cord, and an LCD digital display. When the tip of the immersion probe is placed in a solution, accurate temperature readings appear within seconds. The digital display eliminates the possibility of error that can result from the misreading of stem-type dial thermometers. Detachable immersion probes are time- savers for the busy x-ray department that has several film processors. A permanently positioned probe eliminates the need for removing the cover of the processor. Simply leave one probe in each processor and plug it into the instrument. The use of multiple probes, each remaining in a specific tank, also eliminates possible cross-contamination of chemicals from tank to tank. When mounting the probe permanently to the processor, the probe tip should be positioned so it is approximately 2” beneath the liquid’s surface. Make sure the probe housing is out of the solution.

This battery-operated, shock-resistant read-out unit has solid-state electronics and needs no adjustments to maintain high accuracy. Calibration is traceable to the National Institute of Standards and Technology.

This instrument comes with a high-impact plastic storage case with room for additional probes.

NOTE

This unit is calibrated to standards traceable to the National Institute of Standard and Technology (NIST)

1-1

Image 5
Contents Nuclear Associates Portable Digital ThermometerOperators Manual August Manual No. 07-402-1 RevBlank Page Operating Instructions Table of ContentsIntroduction Blank Page 1.1 Introduction IntroductionSection IntroductionOperators Manual Nuclear Associates2.1 Measure Temperature Section Operating Instructions2.2 Display Min/Max Readings 2.3 Clear Min/Max Memory2.8 Battery Replacement 2.7 Disable the Auto-Shutoff2.9 Specifications 2.6 Backlight the DisplayBlank Page 6045 Cochran Road Cleveland, Ohio 44139 Fluke Biomedical