EMC Tables

Guidance and manufacturer's declaration-electromagnetic immunity

The UT-302PBT Thermometer is intended for use in the electromagnetic environment specified below.The customer or the user of the UT-302PBT Thermometer should assure that it is used in such an environment.

Immunity

IEC 60601 test

Compliance

Electromagnetic environment-guidance

test

level

level

 

Radiated

3 V/m

3 V/m

Portable and mobile RF communications equipment

RF IEC

80 MHz to 2,5

 

should be used no closer to any part of the UT-302PBT

61000-4-3

GHz

 

Thermometer , including cables, than the

 

 

 

recommended separation distance calculated from

 

 

 

the equation applicable to the frequency of the

 

 

 

transmitter.

 

 

 

Recommended separation distance

 

 

 

d=1.2 √ P

 

 

 

d=1.2 √ P 80 MHz to 800 MHZ

 

 

 

d=2.3 √ P 800 MHz to 2,5 Ghz

 

 

 

where P is the maximum output power rating of the

 

 

 

transmitter in watts (W) according to the transmitter

 

 

 

manufacturer and d is the recommended separation

 

 

 

distance in metres (m).

 

 

 

Field strengths from fixed RF transmitters, as

 

 

 

determined by an electromagnetic site survey-a,

 

 

 

should be less than the compliance level in each

 

 

 

frequency range-b.

 

 

 

Interference may occur in the vicinity of equipment

 

 

 

marked with the following symbol:

NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the UT-302PBT Thermometer is used exceeds the applicable RF compliance level above, the UT-302PBT Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the UT-302PBT Thermometer.

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

19

Image 21
Contents Digital Bluetooth Therometer Index Introduction Important Information Before Use Description of LCD Display Product IdentificationLow battery warning Battery InstallationReplacing the Battery Battery-operatedChanging Modes C/F Switchable Tips for Measuring Human Temperature Measuring Body Temperature in the Ear Measuring Body Temperature in the Ear Measuring Body Temperature in the Ear Readings Memory Recall Memory FunctionOperating The Bluetooth Function What You Need Bluetooth LEDCommunications Interface Applied Standards Care and MaintenanceProbe and Tip ThermometerTechnical Specifications Error CodesFCC Statement FCC Statement Cispr EMC TablesRF IEC EMC Tables Page ENGINEERING, INC