MODEL 315 HUMIDITY TEMPERATURE METER
5.2 Humidity Calibration |
1. Turn the unit off. Press and hold the HOLD and °C/°F buttons. While |
these two buttons are pressed, turn the power on. Release all buttons |
MODEL 315 HUMIDITY TEMPERATURE METER
3.0 Symbol Definition and Button Locations:
| and all the segments on the display will blink. |
2. | Within 3 seconds of releasing these buttons, press the UREL button to |
| enter the calibration mode, or the instrument will revert to normal |
| operating mode. |
3. | Upon entering the calibration mode, ‘CAL1’ will appear on the second |
| display. |
4. | Place the probe in a stable 32.8% RH environment at 25°C and allow |
| the system to stabilize for 30 minutes. Press the MAX MIN button to |
| initiate calibration. A single long beep will indicate calibration is |
| complete, and the instrument will advance into the ‘CAL2’ mode, as |
| indicated on the second display. |
5. | Insert the probe in a stable 75.3% RH environment at 25°C and again |
Data Hold indication
Minmum indication
Maximum indication
Low battery indication ![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
![]()
Auto power off
indication
Relative indication Humidity display
Unit indication
Temperature display
allow the system to stabilize for 30 minutes. Press the MAX MIN button |
to initiate calibration. A single long beep will indicate calibration is |
complete, and the instrument will revert to normal operating mode. |
5.3 Restoring Factory Calibration |
1. To restore factory default calibration values, turn the instrument off. |
Press the HOLD and °C/°F buttons. While these two buttons are |
pressed, turn the power on. Release all buttons and all the segments |
1
Feature Locations:
1) Dust mask
2) LCD display
3) ON/OFF button
4) MAX MIN button
5) HOLD button
6) RELATIVE button
7) °C, °F control button
on the display will blink. | |
2. Within | 3 seconds of releasing these buttons, again press the HOLD |
button | to restore factory calibration values. A single long beep will |
indicate restoration is complete. | |
HUMIDITY
8) Battery cover
2 | 9V BATTERY |
| NEDA 1604 6F22 006P |
3
4MAX MIN
5HOLD
6 | REL |
|
7 |
| OPEN |
¢XC / ¢XF | 8 | |
TEMP: | 20 C | 60 C |
| 4 F | 140 F |
HUMIDITY: 0 | 100 %RH | |
5 | 2 |