HH806U/HH806AU
MULTILOGGER THERMOMETER
M4502/0707
Shop online at: omega.com
For latest product manuals: omegamanual.info
OMEGAnet→ | Internet |
omega.com | info@omega.com |
Servicing North America:
U.S.A: | One Omega Drive, Box 4047 | |
| Stamford, CT |
|
ISO 9001 | Tel: (203) | FAX: (203) |
Certified |
| |
Canada: | 976 Bergar |
|
| Laval (Quebec) H7L 5A1, Canada | |
| Tel: (514) | FAX: (514) |
|
|
For immediate technical or application assistance:
U.S.A | Sales Service: |
and | Customer Service: |
Canada: | Engineering Service: |
| TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA |
Mexico: | En Espanol: (001) |
| |
| FAX: (001) |
Servicing Europe:
Benelux: | Postbus 8034, 1180 LA Amstelveen, The Netherlands | |
| Tel: +31 (0)20 3472121 | FAX: +31 (0)20 6434643 |
| Toll Free in Benelux: 0800 0993344 | |
|
| |
Czech | Frystatska 184, 733 01 Karviná, Czech Republic | |
Republic: | Tel: +420 (0)59 6311899 | FAX: +420 (0)59 6311114 |
| Toll Free: | |
France: | 11, rue Jacques Cartier, 78280 Guyancourt, France | |
| Tel: +33 (0)1 61 37 2900 | FAX: +33 (0)1 30 57 5427 |
| Toll Free in France: 0800 466 342 | |
|
| |
Germany/ | Daimlerstrasse 26, | |
Austria: | Tel: + 49 (0)7056 | FAX: +49 (0)7056 |
| Toll Free in Germany: 0800 639 7678 | |
|
| |
United | One Omega Drive, River Bend Technology Centre | |
Kingdom: | Northbank, Irlam, Manchester | |
ISO 9002 | M44 5BD United Kingdom |
|
Certified | Tel: +44 (0)161 777 6611 | FAX: +44 (0)161 777 6622 |
| Toll Free in United Kingdom: | |
|
|
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, human applications.
INTRODUCTION
The instrument is a portable digital thermometer that measures external thermocouples of type K, J, R, S, T, E, N. The thermocouples types comply with the N.I.S.T. – ITS 90 standard reference temperature/voltage tables. The thermometer features a dual thermocouple input, an adjustable T/C offset and an USB interface for uploading data to a PC using optional software and cable.
SAFETY INFORMATION
WARNING
To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature
measurement in microwave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple leads. To prolong lead life, avoid sharp bends in the leads, especially near the connector.
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit
Measurement Range:
*Based on the
Accuracy:
K/J/T/E-TYPE
±(0.05% rdg + 0.3°C) on
N-TYPE
±(0.05% rdg + 0.8°C) on
±(0.05% rdg + 2°C) on 0°C to 1767°C ±(0.05% rdg + 4°F) on 32°F to 3212°F
Temperature Coefficient:
0.1times the applicable accuracy specification per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to122°F).
Input Protection:
24V dc or 24V ac rms maximum input voltage on any combination of input pins.
Reading Rate: 2.5 time per second.
ENVIRONMENTAL
Ambient Operating Ranges:
0°C to 50°C (32°F to 122°F) <80% R.H.
Storage Temperature:
GENERAL
Display:
There are three display areas on the HH806 series LCD (liquid crystal display). The Main and Second displays are 4 ½ digits with a maximum reading of 19999. These are used for displaying the value of T1 or T2. The Third display is for the date, time, or the differential value of T1 to T2.
Overload:
Battery: 1.5V x 4 PCS (SIZE AAA)
Battery Life: 120 hours typical with carbon zinc battery.
Dimensions: 160mm(H) x 83mm(W) x 38mm(D).
Weight: Approx. 365g including batteries.
Supplied Thermocouple:
1 meter (40”) type K insulated beaded wire thermocouple. Maximum insulation temperature is 482°C (900°F). Thermocouple accuracy is ±1.1°C or 0.4% of reading (whichever is greater) from 0°C to 1250°C.
Wire Communication Protocol:
2400 baud rate. (HH806U)
19200 baud rate. (HH806A)
External Connections: | 1 | 2 |
1. USB Port