HH 502/HH 503 CALIBRATION PROCEDURE

Note: The following calibration procedure should perform only by qualified technicians who have access to the items as following.

Equipment: The class of calibrator had better 10 times greater than the measured meter.

1.Set the jumper position on J1 position.

2.The input selection change to T1 position, then input DCV 60mV after the display is stabilized and press “ ENTER “ key.

3.Input 0￿F after the display is stabilized and press “ ENTER “ key to complete calibration.

4.After adjustment set the jumper back to J3 position.

1

Image 12
Contents Omega Omega.comTM Safety Information IntroductionSpecifications ElectricalEnvironmental Power Switch F Selecting the Temperature ScaleHold Mode only Main display Operating InstructionsMIN MAX with Time record Mode J T1 Input Thermocouple Type SelectREL Relative mode Sec. Minu. Selecting the Time scale Relative value Recall display mode 11 Hi/Lo Limits value Recall display mode12 Hi/Lo Limits mode only Main display Operator Maintenance Battery ReplacementHH 502/HH 503 Calibration Procedure Warranty / Disclaimer Temperature