Troubleshooting and Displays
The thermometer has been designed to display feedback/warnings to ensure proper usage and accurate temperature measurement. When an error or indication appears, one “beep” signal will be heard once every 5 seconds for 10 seconds and then the thermometer will turn off automatically. In the event of a malfunction or incorrect temperature measurement, the following warnings will be displayed:
Display | Meaning | Action |
| The measured | Turn the unit OFF and then ON |
| temperature is below | again. Retake temperature as |
| 95°F (35°C). | described in “Using Your |
|
| Thermometer” section. |
| The measured temperature | Turn the unit OFF and then ON |
| is above 107.6°F (42°C). | again. To verify the reading, retake |
| Or thermometer was | temperature as described in “Using |
| placed in body position | Your Thermometer” section. |
| before two beeps were |
|
| heard. |
|
| The thermometer is too | Verify room temperature is within |
| cold or too hot to | 50°F to 104°F (16°C to 40°C). |
| operate. | Retake temperature as described in |
|
| “Using Your Thermometer” section. |
| Temperature did not | Turn the unit OFF, wait two |
| register because of poor | minutes, turn it ON again and |
| contact, patient movement | retake temperature as described in |
| or poor placement. | “Using Your Thermometer” section. |
| Unit malfunctioned. | Do not use the thermometer – |
|
| contact distributor for instructions. |
| Battery icon is on | The batteries have enough |
| continuously. | remaining life for a few more |
|
| measurements. Replace batteries. |
| Only battery icon is on | The batteries are drained - replace |
| and the device | batteries (see “Replacing Batteries” |
| automatically turns | section). |
| itself OFF. |
|
|