4. Replace the battery cover.

13

24

IMPORTANT: Only use a CR1225 3V Lithium battery. Use of another type of battery may present a risk of fire. The battery may explode if it is handled improperly. Do not recharge, disassemble, heat the battery above 212˚F/100˚C, or incinerate. Dispose of all used batteries in the appropriate way.

Troubleshooting

Display

Meaning

Action

 

Battery voltage is

Replace old battery

 

low.

with a new one

 

 

immediately.

There is no

The thermometer

Reset the

display or there

should be reset or

thermometer by

is a fixed

the battery should

removing and

display showing

be replaced.

reinstalling the same

Er0-Er7

 

battery; replace the

 

 

battery with a new

 

 

one; call BébéSounds®

 

 

service center.

 

Ambient

Place thermometer in

 

temperature is out

a room within the

 

of range (50.0-

temperature range

 

104.0˚F/10.0-

and wait about 5

 

43.0˚C).

minutes before taking

 

 

a new reading.

 

Unstable

Take a new

 

temperature

measurement while

 

reading.

holding the

 

 

thermometer in a

 

 

stable position.

 

Temperature is

Take a new reading

 

outside

and wait for the beep

 

measurement

before removing the

 

range (86.0-

thermometer.

 

109.4˚F/30.0-

 

 

43.0˚C).

 

 

Ambient

Follow Step 6 in How

 

temperature is

to Use Section.

 

over 86˚F/30˚C.

 

If you continue to have difficulties obtaining a reading call 1-888-232-6476.

Technical Specifications

For home use only

Temperature readings are to the nearest 0.01˚F/C

Temperature measurement range:

86.0– 109.4˚F/30.0 – 43.0˚C

Operating temperature range:

50.0– 104.0˚F/10.0 – 43.0˚C

Requires 1 CR1225 3V Lithium battery (included)

Applied Standards and Approvals

Meets American Society for Testing and Materials (ASTM) standard E1112-98 (Accuracy)

Meets Japanese Standards Association (JIS) standard JIS T 1140:1998 (Accuracy)

Meets EMC Standard EN60601-1 (Safety)

Limited Product Warranty

The BébéSounds® 5-Second 3-Way Flexible Tip Thermometer, distributed by Unisar Inc., has a limited warranty against defects for a period of 2 years from the original purchase date. Battery life or damage due to misuse or abuse are excluded from this warranty.

In-Warranty Procedure

In the event service is required due to defect or malfunction during the warranty period, Unisar Inc. will replace or repair the product under warranty. The repair or replacement of the defective item will be free of charge. Repair or replacement of the defective item is subject to verification of the malfunction or defect when delivered to:

Unisar® Service Center

15 West 36th Street

New York, NY 10018

Please enclose dated proof of purchase with your return.

This thermometer is not a toy. Do not allow children to play with it. Keep out of reach of children when not in use.

6

7

Image 4
Contents Lundi-Vendredi de 9 heures à 17 heures HNE De Lunes a Viernes de 900 a 1700 hora del esteUnisar Inc West 36th Street New York, NY Visit us atOral, Rectal, and Axillary underarm Temperature Ranges How to UseIntroduction InstructionsBattery Installation For Oral MeasurementFor Rectal Measurement For Axillary Underarm MeasurementTroubleshooting Technical SpecificationsLimited Product Warranty In-Warranty ProcedureColor Portable Video & Sound Monitor Movement Sensor with Sound MonitorPortable miniFRIDGE TV ListenerIntroducción Rangos de Temperatura Oral, Rectal y Axilar bajo el brazoInstrucciones Para Usar Por Primera VezPara Medición Oral Instalación de la BateríaPara Medición Rectal Solución de Problemas Especificaciones TécnicasGarantía Limitada del Producto Procedimiento En Garantía