MODEL 140,141, and 142 sensors

INSTALLATION

INSTALLATION – 141 SENSOR

1.Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee.

2.Remove the plastic shipping cap from the sensor.

3.Screw the sensor into the fitting. Use pipe tape on the threads. See Figure 3.

DWG. NO.

40014204

REV.

B

FIGURE 3. Installing model 141 sensor

INSTALLATION – 142 SENSOR

1.Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee. See Figure 4.

2.Remove the plastic shipping cap from the sensor.

3.Screw the sensor into the fitting. Use pipe tape on the threads. DO NOT tighten the sensor compression fitting until the sensor is correctly positioned.

4.If necessary, loosen the sensor compression fitting and position the sensor so that the tip of the sensor is at least 1-inch (25 mm) from the far wall of the pipe.

5.Tighten the compression fitting using the procedure shown in Figure 2.

 

 

 

 

 

DWG. NO.

REV.

 

 

40014204

B

 

 

 

 

 

 

 

 

 

FIGURE 4. Installing model 142 sensor

3

Image 3
Contents Conductivity Sensors Sensor SpecificationsInstallation Installation 140 Sensor with Ball Valve KIT PNSensor Section on Installation 142 Sensor Installation 141 SensorWiring Wire Color and Connections in SensorWiring Diagrams Model Xmt-C-10 Wiring Panel System pressure must be less than 100 psig 791 kPa Retracting and Inserting the Model 140 SensorRetracting the Sensor Inserting the SensorStand clear of the sensor Removing and Reinstalling the Model 142 SensorReinstalling the Sensor Removing the SensorCalibration Replacing Model 140 Sensor Seal MaintenanceCleaning the Sensor Checking Model 140 Retraction RestraintDescription Replacement PartsProblem Probable Cause Solution TroubleshootingSpecifications subject to change without notice